Combined Motion Estimation and Reconstruction in Tomography
نویسندگان
چکیده
If objects or patients move during a CT scan, reconstructions suffer from severe motion artifacts. Time dependent computed tomography (4DCT) tries to minimize these artifacts by estimating motion and/or reconstruction simultaneously. Most current methods assume a known deformation or a reconstruction without artifacts at a certain time point. This work explores the possibilities of estimating the motion model and reconstruction simultaneously. It does so by modifying the simultaneous iterative reconstruction technique (SIRT) to incorporate motion (trans-SIRT) and uses this method in an optimization routine that computes motion and reconstruction at the same time. Results show that the optimization routine is able to estimate motion accurately, assuming only the type of parametrization for the motion model. Our approach can potentially be extended to more complex motion models.
منابع مشابه
Artifact reduction techniques in Cone Beam Computed Tomography (CBCT) imaging modality
Introduction: Cone beam computed tomography (CBCT) was introduced and became more common based on its low cost, fast image procedure rate and low radiation dose compared to CT. This imaging modality improved diagnostic and treatment-planning procedures by providing three-dimensional information with greatly reduced level of radiation dose compared to 2D dental imaging modalitie...
متن کاملRespiratory motion correction in prostate cancer positron emission tomography: A study on patients and phantom simulation
Introduction: To investigate the effects of breathing cycle and tree diaphragm motions on prostate cancer tumors standard uptake value (SUV) during positron emission tomography (PET) and to correct it. Materials and methods: Respiratory motion traces were simulated on the common patient breathing cycle and tree diaphragm motio...
متن کاملClinical Data Evaluation of C-arm-based Motion Compensated Coronary Artery Reconstruction
Angiographic C-arm systems are capable of performing computed tomography (CT) imaging for assisting in therapy planning, performance and assessment. Due to the long acquisition time, C-arm CT of dynamic structures is challenging. Cardiac motion has to be estimated and compensated in the reconstruction step. The quality of the motion estimation algorithm mainly dictates the resulting image quali...
متن کاملSimultaneous Reconstruction and 3d Motion
of Dissertation Presented to the Graduate School of the University of Florida in Partial Fulfillment of the Requirements for the Degree of Doctor of Philosophy SIMULTANEOUS RECONSTRUCTION AND 3D MOTION ESTIMATION FOR GATED MYOCARDIAL EMISSION TOMOGRAPHY By Zixiong Cao December 2003 Chair: David Gilland Major Department: Biomedical Engineering A new method for simultaneous image reconstruction a...
متن کاملMotion Compensated Fan-Beam CT by Enforcing Fourier Properties of the Sinogram
In computed tomography involuntary patient motion can lead to a severe degradation of image quality. Most of the motion estimation methods rely on additional information, such as fiducial markers or an ECG signal. In contrast, data driven motion estimation exists which aims to estimate the motion directly from the acquired projections. This is typically achieved by the optimization of an error ...
متن کامل