Bilateral fatigue fracture of the femoral components in a cruciate-retaining cementless total knee prosthesis.
نویسندگان
چکیده
This article reports a case of bilateral fatigue fracture of the femoral components in a cruciate-retaining uncemented total knee arthroplasty (TKA). A 75-year-old woman (height, 158 cm; weight, 72 kg; body mass index, 29.2) had undergone one-stage bilateral TKA for osteoarthritis 11 years previously at the author's institution. Surgery was performed using an uncemented Flexible Nichidai Knee. Equal tension of the collateral ligaments and normal mechanical axis were achieved during the primary procedure. The patient was an ardent lover of the game of badminton and had higher activity levels with daily playing. At 8 years postoperatively, she started complaining of mild pain in both knees. The pain gradually increased, and at 11 years postoperatively, she had difficulty walking. Anteroposterior radiographs showed narrowing of the medial joint space, indicating wear of the polyethylene insert. Lateral radiographs showed signs of broken implants in both knees. There were no signs of gross implant loosening or osteolysis. One-stage revision surgery was performed, and the knees were converted to cemented posterior-stabilized TKAs. At revision, the bilateral femoral components were found to be fractured at the junction between the trochlear flange and the medial condyle, anteriorly to the medial peg. The polyethylene insert showed mild wear at the medial middle portion. In the majority of case reports, stress fractures of the femoral component have predominantly affected the medial condyle, following uncemented implantation of fixed-bearing knees. In this case, failure of bone ingrowth in uncemented components, higher body mass index, and a higher athletic activity led to fatigue fracture of the femoral components.
منابع مشابه
A rare case of bilateral large osteolysis following cemented and cementless total knee arthroplasties
Open Access This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the source is credited. DOI 10.3109/17453674.2013.752693 A 66-year old woman (BMI 36) presented for routine followup after undergoing total knee arthroplasty (TKA) in the right knee 12 years ...
متن کاملNontraumatic Fracture of the Femoral Condylar Prosthesis in a Total Knee Arthroplasty Leading to Mechanical Failure
This paper reports a case of fatigue fracture of the femoral component in a cruciate-retaining cemented total knee arthroplasty (TKA). A 64-year-old man had undergone a primary TKA for osteoarthritis 10 years previously at another institution using the PFC-Sigma prosthesis. The patient recovered fully and was back to his regular activities. He presented with a history of sudden onset pain and l...
متن کاملPosterior cruciate-retaining, rotating-platform total knee arthroplasty: minimum 4-year follow-up study.
Rotating-platform, mobile-bearing total knee arthroplasty was initially developed using a posterior cruciate-sacrificing technique and design. Posterior cruciate-substituting and -retaining, rotating-platform total knee arthroplasty designs were developed later. The purpose of this study was to evaluate a minimum 4-year follow-up of total knee arthroplasty performed with a posterior cruciate-re...
متن کاملClinical results of Hi-tech Knee II total knee arthroplasty in patients with rheumatoid athritis: 5- to 12-year follow-up
BACKGROUND Total knee arthroplasty (TKA) is a common form of treatment to relieve pain and improve function in cases of rheumatoid arthritis (RA). Good clinical outcomes have been reported with a variety of TKA prostheses. The cementless Hi-Tech Knee II cruciate-retaining (CR)-type prosthesis, which has 6 fins at the anterior of the femoral component, posterior cruciate ligament (PCL) retention...
متن کاملPolyethylene Bearing Stress Fractures after Mobile-bearing Cruciate-retaining Total Knee Arthroplasty: Case Reports
Some reported bearing fractures are due to entrapment of the bearing between the femoral component and posterior edge of the tibial base plate in unstable meniscal –bearing design TKA. However, so far no cases of bearing fracture has been reported for the rotating platform or anteroposterior gliding designs of mobile bearing TKA. The authors describe two cases of polyethylene stress fracture af...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Orthopedics
دوره 34 10 شماره
صفحات -
تاریخ انتشار 2011