Bench evaluation of 7 home-care ventilators.

نویسندگان

  • Thomas C Blakeman
  • Dario Rodriquez
  • Dennis Hanseman
  • Richard D Branson
چکیده

BACKGROUND Portable ventilators continue to decrease in size while increasing in performance. We bench-tested the triggering, battery duration, and tidal volume (V(T)) of 7 portable ventilators: LTV 1000, LTV 1200, Puritan Bennett 540, Trilogy, Vela, iVent 101, and HT50. METHODS We tested triggering with a modified dual-chamber test lung to simulate spontaneous breathing with weak, normal, and strong inspiratory effort. We measured battery duration by fully charging the battery and operating the ventilator with a V(T) of 500 mL, a respiratory rate of 20 breaths/min, and PEEP of 5 cm H(2)O until breath-delivery ceased. We tested V(T) accuracy with pediatric ventilation scenarios (V(T) 50 mL or 100 mL, respiratory rate 50 breaths/min, inspiratory time 0.3 s, and PEEP 5 cm H(2)O) and an adult ventilation scenario (V(T) 400 mL, respiratory rate 30 breaths/min, inspiratory time 0.5 s, and PEEP 5 cm H(2)O). We measured and analyzed airway pressure, volume, and flow signals. RESULTS At the adult settings the measured V(T) range was 362-426 mL. On the pediatric settings the measured V(T) range was 51-182 mL at the set V(T) of 50 mL, and 90-141 mL at the set V(T) of 100 mL. The V(T) delivered by the Vela at both the 50 mL and 100 mL, and by the HT50 at 100 mL, did not meet the American Society for Testing and Materials standard for V(T) accuracy. Triggering response and battery duration ranged widely among the tested ventilators. CONCLUSIONS There was wide variability in battery duration and triggering sensitivity. Five of the ventilators performed adequately in V(T) delivery across several settings. The combination of high respiratory rate and low V(T) presented problems for 2 of the ventilators.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Near-infrared spectroscopy with vascular occlusion test may not be the adequate tool to explore microcirculation in pulmonary arterial hypertension.

essentials of respiratory care, 4th edition. St. Louis:Mosby Elsevier; 2005:103. 13. L’Her E, Roy A. Bench tests of simple, handy ventilators for pandemics: performance, autonomy, and ergonomy. Respir Care 2011;56(6):751-760. 14. Lyazidi A, Thille AW, Carteaux G, Galia F, Brochard L, Richard JC. Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventila...

متن کامل

Evaluation of ventilators for mouthpiece ventilation in neuromuscular disease.

BACKGROUND Daytime mouthpiece ventilation is a useful adjunct to nocturnal noninvasive ventilation (NIV) in patients with neuromuscular disease. The aims of the study were to analyze the practice of mouthpiece ventilation and to evaluate the performance of ventilators for mouthpiece ventilation. METHODS Practice of mouthpiece ventilation was assessed by a questionnaire, and the performance of...

متن کامل

Performance of ventilators for noninvasive positive-pressure ventilation in children.

The aim of the present study was to evaluate the performance characteristics of all the ventilators proposed for home noninvasive positive-pressure ventilation in children in France. The ventilators (one volume-targeted, 12 pressure-targeted and four dual) were evaluated on a bench which simulated six different paediatric ventilatory patterns. For each ventilator, the quality of the inspiratory...

متن کامل

Effect of leak and breathing pattern on the accuracy of tidal volume estimation by commercial home ventilators: a bench study.

BACKGROUND New home ventilators are able to provide clinicians data of interest through built-in software. Monitoring of tidal volume (VT) is a key point in the assessment of the efficacy of home mechanical ventilation. OBJECTIVE To assess the reliability of the VT provided by 5 ventilators in a bench test. METHODS Five commercial ventilators from 4 different manufacturers were tested in pr...

متن کامل

Bench tests of simple, handy ventilators for pandemics: performance, autonomy, and ergonomy.

BACKGROUND It has been pointed out that in the wake of a virulent flu strain, patients with survivable illness will die from lack of resources unless more ventilators are made available. Numerous disaster-type ventilators are available, but few evaluations have been performed. OBJECTIVE To compare simple, lightweight, and handy ventilators that could be used in the initial care of patients wi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Respiratory care

دوره 56 11  شماره 

صفحات  -

تاریخ انتشار 2011