False positive rate of carbon monoxide saturation by pulse oximetry of emergency department patients.

نویسندگان

  • Lindell K Weaver
  • Susan K Churchill
  • Kayla Deru
  • Darryl Cooney
چکیده

BACKGROUND Symptoms of carbon monoxide (CO) poisoning are non-specific. Diagnosis requires suspicion of exposure, confirmed by measuring ambient CO levels or carboxyhemoglobin (COHb). An FDA-approved pulse oximeter (Rad-57) can measure CO saturation (S(pCO)). The device accuracy has implications for clinical decision-making. METHODS From April 1 to August 15, 2008, study personnel measured S(pCO) and documented demographic factors at time of clinical blood draw, in a convenience sample of 1,363 subjects presenting to the emergency department at Intermountain Medical Center, Murray, Utah. The technician then assayed COHb. COHb and S(pCO) values were compared by subject; false positive or negative values were defined as S(pCO) at least 3 percentage points greater or less than COHb level, reported by the manufacturer to be ± 1 SD in performance. RESULTS In 1,363 subjects, 613 (45%) were male, 1,141 (84%) were light-skinned, 14 in shock, 4 with CO poisoning, and 122 (9%) met the criteria for a false positive value (range 3-19 percentage points), while 247 (18%) met the criteria for a false negative value (-13 to -3 percentage points). Risks for a false positive S(pCO) reading included being female and having a lower perfusion index. Methemoglobin, body temperature, and blood pressure also appear to influence the S(pCO) accuracy. There was variability among monitors, possibly related to technician technique, as rotation of monitors among technicians was not enforced. CONCLUSIONS While the Rad-57 pulse oximeter functioned within the manufacturer's specifications, clinicians using the Rad-57 should expect some S(pCO) readings to be significantly higher or lower than COHb measurements, and should not use S(pCO) to direct triage or patient management. An elevated S(pCO) could broaden the diagnosis of CO poisoning in patients with non-specific symptoms. However, a negative S(pCO) level in patients suspected of having CO poisoning should never rule out CO poisoning, and should always be confirmed by COHb.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Non-invasive carboxyhemoglobin monitoring: screening emergency medical services patients for carbon monoxide exposure.

INTRODUCTION Carbon monoxide (CO) toxicity is a significant health problem. The use of non-invasive pulse CO-oximetry screening in the emergency department has demonstrated that the rapid screening of numerous individuals for CO toxicity is simple and capable of identifying occult cases of CO toxicity. OBJECTIVE The objective of this study was to extend the use of this handheld device to the ...

متن کامل

Victim of carbon monoxide poisoning identified by carbon monoxide oximetry.

BACKGROUND Carbon monoxide (CO) is known as a leading cause of unintentional poisoning death in many countries. Diagnosis is usually made by measuring carboxyhemoglobin (COHb), but due to the non-specific symptoms, the crucial step is considering CO poisoning. A possible solution might be screening emergency department (ED) patients. A cutaneous sensor that measures oxygen and CO saturation sim...

متن کامل

Accuracy of noninvasive multiwave pulse oximetry compared with carboxyhemoglobin from blood gas analysis in unselected emergency department patients.

STUDY OBJECTIVE Accurate and timely diagnosis of carbon monoxide (CO) poisoning is difficult because of nonspecific symptoms. Multiwave pulse oximetry might facilitate the screening for occult poisoning by noninvasive measurement of carboxyhemoglobin (COHb), but its reliability is still unknown. We assess bias and precision of COHb oximetry compared with the criterion standard blood gas analysi...

متن کامل

Reliability of new pulse CO-oximeter in victims of carbon monoxide poisoning.

STUDY OBJECTIVE The purpose of this study was to evaluate the reliability of noninvasive real-time measurement of carboxyhemoglobin (COHb) using a pulse CO-oximeter in victims of carbon monoxide poisoning (COP). METHODS During the 7-month study period, pulse CO-oximetry was measured on patients admitted to the emergency department (ED) for suspected COP. Each patient included in the study und...

متن کامل

Pulse oximetry in the accident and emergency department.

The early detection and treatment of hypoxia is important to reduce patient morbidity in the accident and emergency department. At present, the commonly used methods all have practical difficulties in the urgent situations that prevail in the department. Pulse oximetry has recently become a method of choice in many anaesthetic and intensive care units for the continuous monitoring of oxygenatio...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Respiratory care

دوره 58 2  شماره 

صفحات  -

تاریخ انتشار 2013