Separation of Signals Acquired with EMAT in Dual Coil Configuration
نویسندگان
چکیده
In this paper, the proposal of electromagnetic acoustic transducer (EMAT) in dual coil configuration with consequent application of signal de-noising algorithm is presented in detail. Dual EMAT contains two operated coils in a single housing. Both coils transmit and receive the ultrasonic signal. Based on the proposed coil configuration, two signals characterizing material structure at the same place are received and consequently evaluated. This configuration makes the flaw detection process more accurate and the noise level can be efficiently suppressed using advanced signal processing algorithms. In this study, we also investigate the proposal of signal processing algorithm used for signal and noise separation. The proposed algorithm based on independent component analysis (ICA) uses the statistical information to separate useful signal as fault echo from the undesirable echoes that are caused by the scattering from grains contained in materials. Using our proposed algorithm it is possible to recognize the fault echo that is hidden under the noise level from backscattering noise. The performance of both, proposed EMAT in dual coil configuration and algorithm is compared on simulated and acquired EMAT signals.
منابع مشابه
Effect of Coil Configuration on Conversion Efficiency of EMAT on 7050 Aluminum Alloy
Compared to traditional ultrasonic detection methods, the electromagnetic acoustic transducer (EMAT) technique can be applied in many hostile environments such as elevated temperatures, on-line inspections, etc. However, the EMAT technique has a low conversion efficiency. This paper develops a numerical model to study the effects of the coil configurations on the conversion efficiency of shear-...
متن کاملAn Improved Design of the Spiral-Coil EMAT for Enhancing the Signal Amplitude
The low energy transition efficiency of electromagnetic ultrasonic transducer (EMAT) is a common problem in practical application. For the purpose of enhancing the amplitude of the received signal, an improved double-coil bulk wave EMAT is proposed for the thickness measurement of metallic block. This new configuration of magnets consists of a solid cylindrical magnet and a ring-shaped magnet e...
متن کاملApplication of Pulse Compression Signal Processing Techniques to Electromagnetic Acoustic Transducers (emats) for Non-contact Thickness Measurements and Imaging
A pair of non-contact EMATs has been used for thickness measurements and imaging of metallic plates. This was performed using wide bandwidth EMATs and pulsecompression signal processing techniques, using chirp excitation. This gives a greatly improved Signal to Noise Ratio (SNR) for non-contact experiments, increasing the speed of data acquisition. A numerical simulation of the technique has co...
متن کاملThe Optimization of Lamb and Rayleigh Wave Generation Using Wideband-low-frequency Emats
This paper describes a non-contact ElectroMagnetic Acoustic Transducer (EMAT) that can be used to generate both Lamb and Rayleigh waves on metal samples. The generated waves are wideband and low frequency with a dominant frequency content centred on approximately 200kHz extending to around 500kHz. Detection of the waves is achieved using a linear coil detection EMAT. The transducers (generator ...
متن کاملA New Omni-Directional EMAT for Ultrasonic Lamb Wave Tomography Imaging of Metallic Plate Defects
This paper proposes a new omni-directional electromagnetic acoustic transducer (EMAT) for the ultrasonic Lamb wave (ULW) tomography imaging (TI) of defects in metallic plates. The proposed EMAT is composed of a permanent magnet and a coil with a contra-flexure structure. This new EMAT coil structure is used for omni-directional ULW transmission and reception and ULW TI for the first time. The t...
متن کامل