Dynamic contour tonometry in comparison to intracameral IOP measurements.

نویسندگان

  • Andreas G Boehm
  • Anja Weber
  • Lutz E Pillunat
  • Rainer Koch
  • Eberhard Spoerl
چکیده

PURPOSE The purpose of the study was to examine the effect of central corneal thickness (CCT), corneal curvature, astigmatism, axial length (AL), and age on measurements with the Pascal Dynamic Contour Tonometer (DCT). METHODS In a prospective clinical trial 75 eyes of 75 patients undergoing phacoemulsification were examined. Before phacoemulsification, the anterior chamber was cannulated at the temporal corneal limbus. In a closed system the IOP was directly set to 15, 20, or 35 mm Hg with a manometric water column. IOP measurements taken by DCT were compared to intracameral measurements with a precision reference pressure sensor. RESULTS Measurements from 60 patients were suitable for statistical analysis. At IOP of 15 mm Hg, the mean difference between IOP measured by DCT and intracameral IOP was -0.02 +/- 1.32 mm Hg; at 20 mm Hg it was -0.2 +/- 1.44 mm Hg and at 35 mm Hg, -0.84 +/- 1.90 mm Hg. The concordance coefficient according to Lin was 0.9763, showing good agreement between DCT- and intracamerally measured IOP. There was a statistically significant correlation between the difference in IOP measured by DCT minus intracameral IOP and CCT (P = 0.0291, R2 = 0.00012). All other parameters had no statistically significant effect on the difference between DCT and intracameral IOP (corneal curvature, P = 0.6094, R2 = 0.00367; age, P = 0.9198, R2 = 0.000003; astigmatism, P = 0.1564, R2 = 0.08497; and axial length, P = 0.9484, R2 = 0.00008). CONCLUSIONS Measurements with the DCT showed good concordance with intracameral IOP. CCT exerted a statistically significant but clinically irrelevant effect on measurements with the DCT.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dynamic Contour Tonometry, Tono-Pen XL®, and Goldmann Applanation Tonometry in Comparison to Intracameral Intraocular Pressure (IOP) Measurements in Patients with Corneal Pathologies

Unspecified Posted at the Zurich Open Repository and Archive, University of Zurich ZORA URL: http://doi.org/10.5167/uzh-101903 Published Version Originally published at: Neuburger, M; Grosswendt, J; Lautebach, S; Maier, P; Birnbaum, F; Böhringer, D; Funk, J; Jordan, J F; Reinhard, T (2014). Dynamic Contour Tonometry, Tono-Pen XL®, and Goldmann Applanation Tonometry in Comparison to Intracameral...

متن کامل

Dynamic contour tonometry: a comparative study on human cadaver eyes.

OBJECTIVE To compare intraocular pressure measurements obtained by recently introduced dynamic contour tonometry (DCT), Goldmann applanation tonometry (GAT), pneumatonometry (PTG), and intracameral manometry in human cadaver eyes. METHODS Sixteen freshly enucleated human cadaver eyes were deepithelialized and dehydrated with dextran. A tube was placed in the anterior chamber and connected to ...

متن کامل

Goldmann and error correcting tonometry prisms compared to intracameral pressure

BACKGROUND Compare Goldmann applanation tonometer (GAT) prism and correcting applanation tonometry surface (CATS) prism to intracameral intraocular pressure (IOP), in vivo and in vitro. METHODS Pressure transducer intracameral IOP was measured on fifty-eight (58) eyes undergoing cataract surgery and the IOP was modulated manometrically to 10, 20, and 40 mmHg. Simultaneously, IOP was measured ...

متن کامل

Comparison of dynamic contour tonometry with Goldmann applanation tonometry following Descemet's stripping automated endothelial keratoplasty (DSAEK).

BACKGROUND Intraocular pressure (IOP) measurements using Goldmann applanation tonometry (GAT) are influenced by central corneal thickness (CCT) whereas dynamic contour tonometry (DCT) allows for IOP measurements independent of CCT. After Descemet's stripping automated endothelial keratoplasty (DSAEK) the CCT is, compared to healthy eyes, increased. The objective of this prospective study was to...

متن کامل

Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings.

BACKGROUND Several authors reported incorrect high intraocular pressure (IOP) values in eyes with a thick cornea using applanation tonometry. This hypothesis was checked by comparing applanation tonometry with direct intracameral manometry. METHODS 73 patients, scheduled for intraocular surgery, were enrolled. Immediately before surgery, the following were registered: (i) central corneal thic...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Investigative ophthalmology & visual science

دوره 49 6  شماره 

صفحات  -

تاریخ انتشار 2008