Tilted Transceiver Array for Ultra-high Field MRI
نویسندگان
چکیده
Introduction: Ultra-high field transceiver coil array holds many advantages and unique capabilities over the conventional non-array volume transceiver coils such as B1 shimming/manipulating and parallel excitation for human studies. The transceiver array can be either an array of loops or an array of straight microstrip elements [1-3]. Usually the mutual coupling in such arrays is minimized by using the dedicated decoupling network across the adjacent coil elements. The isolation better than -18dB, however, is technically challenging in practice. In addition, the depth of B1 field may not be optimized due to the small gap between coil elements and the RF shield (or ground plane), although a ground plane benefits coil isolation and radiation loss reduction. The volume arrays with tilted microstrip and loop elements are proposed and investigated in this work. In those arrays, each element was tilted with a certain angle for achieving sufficient decoupling without using dedicated decoupling network. 8-ch human knee transceiver arrays with straight microstrips and loops were designed and tested at 7T. Decoupling is significantly improved for both arrays, and the B1 field is also increased in the imaging region for the microstrip array compared with non-tilted case. Methods: The receive-only venetian-blind array was introduced [4] for decoupling and G-map optimization. Our investigation shows that tilted elements are more useful for transceiver array decoupling since the element overlapping is not suggested and is extremely difficult to implement (microstrip). We built an 8ch microstrip transceiver array and an 8-ch shielded loop transceiver array for human knee imaging (see Fig 1) using the tilted decoupling technique. Each microstrip element was built on a 1.2 cm thick Teflon board with the length of 16cm and width of 6cm. In the loop array, each loop is 16cm × 6cm, gap between the loop and the RF shield is 1.2cm. To conveniently adjust the tilted angle of each coil element (Fig 2), specific supporting frames were built with 21cm inner i.d. and 28 cm o.d. The isolation between two adjacent coil elements (unloaded case) was measured on bench with different angle θ. We also estimated the B1 strength to tilted angles of a single loop/ microstrip element at the central point of the volume coil. We used a shielded pickup coil to measure the induced voltages from two perpendicular directions. The combined voltage represented the B1 field strength. We simulated the B1 distribution of the tilted microstrip array and the regular microstrip array (not tilted) at 300MHz with XFDTD (Remcom Inc.). To ignore the impact of mutual coupling, current sources with different phases (0°, 45°, 90°...) were placed at two ends of the eight elements. For fair comparison, the B1 were finally normalized by dividing with the mean value of the background. MR imaging experiments with the tilted microstrip array was performed on a 7T/90cm magnet. This scanner is equipped with two quadrature transmit channels and two T/R switches. To test the transceiver arrays on this system, continuous scans were conducted by connecting two coil elements into the transmit channels each time, and combined all sub-images offline. Phantom images and human knee images from a healthy volunteer were acquired with this array by using gradient echo (GRE) images. The acquisition parameters used were TE/TR= 6.9ms/100ms at 20°flip angle. Results: Bench measurements were shown in Fig 3. After tilting the microstrip elements with 30°, the best isolation of -22dB is observed. It is interesting that 28% more B1 strength (efficiency) is also observed at the center point than that with θ = 0° (i.e. not tilted case). For the loop element, the best isolation occurs at 25°. With the 25° tilting, coil efficiency of the single loop dropped 10% than 0° case. After loading with a human knee, the coupling between adjacent elements in the regular microstrip array (not tilted) was relative strong (-10dB ~ -12dB). The isolation was improved to -19dB ~ -25dB after 30° element tilting. B1 fields from a single microstrip and a single loop are shown in Fig 4. Their different B1 field profiles further reveal their increase/decrease trends after the tilting. Fig 5 shows the simulated B1 fields (unloaded case). With the similar B1 strength at the periphery, the tilted array achieved >20% B1 strength at the central part. Fig 6 shows the initial phantom and knee images acquired using the tilted microstrip array. Conclusions: Tilted transceiver microstrip and loop arrays have been demonstrated at 7T to achieve optimized isolation among resonant elements. For the microstrip array with tilted elements, B1 strength is also increased at the depth. This array is particularly useful for large samples with densely-placed coil elements where element decoupling is usually problematic. The future investigation on this design will focus on the parallel imaging performance, safety issues, and the coil efficiency of the tilted transceiver array. Acknowledgments: This work was supported by NIH grants EB004453 UC Discovery ITLbio04-10148 and QB3 opportunity award. The authors would like to thank Dr. Eddy Boskamp for his valuable suggestions on this work. References: [1] E. Boskamp et al. ISMRM 16 2008 pp1094 [2] E. Boskamp et al. ISMRM 15 2007 pp1046 [3] G. Adriany, et al. MRM 53, 2005, p434445. [4] C. J. Hardy et al. ISMRM 13 2005 pp 677
منابع مشابه
Tilted Microstrip Phased Arrays With Improved Electromagnetic Decoupling for Ultrahigh-Field Magnetic Resonance Imaging
One of the technical challenges in designing a dedicated transceiver radio frequency (RF) array for MR imaging in humans at ultrahigh magnetic fields is how to effectively decouple the resonant elements of the array. In this work, we propose a new approach using tilted microstrip array elements for improving the decoupling performance and potentially parallel imaging capability. To investigate ...
متن کاملA Novel 8-Channel Transceive Volume-Array for a 9.4T Animal Scanner
Introduction This work focuses on the design of a novel 8-element TX/RX volume-array for high field animal MRI. The magnetic field strength of current small animal MRI systems is increasing in search of higher SNR and spectral resolution [1]. Along with the move to higher field strengths, RF coil design for small-animal MRI is advancing concurrently [2]. At ultra high field strengths, however, ...
متن کاملActively Detunable 8-channel Small Animal Transceive Volume Array for 9.4T MRI systems
Introduction Ultra-high field MRI systems for small animals have seen a rapid development over recent years in a continuous effort to enhance SNR and spectral resolution of MRI images. To circumvent the associated high-field B1 inhomogeneity problem, methods such as RF shimming and Transmit SENSE have been proposed [1-2]. To benefit from the advantages of these techniques, dedicated phased-arra...
متن کاملA Stripline-like Coil Element Structure for High Field Phased Array Coils and its Application for a 8-channel 9.4T Small Animal Transceive Array
Introduction Ultra-high magnetic field strength, as widely utilised in small animal MRI systems, has advantages of being able to enhance signal-to-noise ratio (SNR) and gain high spectral resolution [1]. However, the homogeneity of B1 field worsens along with the move to higher RF-frequencies. Methods such as RF shimming and transmit SENSE, which require transceive volume-arrays had been propos...
متن کاملDetailed Investigation of Ultra-Wideband Impulse Radio Communication Systems with LeCroy Oscilloscopes
I. OVERVIEW Target of the work is the evaluation of the potential of impulse radio (IR) communication systems with respect to commercial usage. Therefore a reconfigurable ultra-wideband (UWB) software defined radio (SDR) transmitter has been developed and the concept is presented in this application note. Besides common SDR concepts the developed IR transceiver employs first order bandpass (BP)...