Reduction of implant RF heating through modification of transmit coil electric field.
نویسندگان
چکیده
In this work, we demonstrate the possibility to modify the electric-field distribution of a radio frequency (RF) coil to generate electric field-free zones in the body without significantly altering the transmit sensitivity. Because implant heating is directly related to the electric-field distribution, implant-friendly RF transmit coils can be obtained by this approach. We propose a linear birdcage transmit coil with a zero electric-field plane as an example of such implant-friendly coils. When the zero electric-field plane coincides with the implant position, implant heating is reduced, as we demonstrated by the phantom experiments. By feeding RF pulses with identical phases and shapes but different amplitudes to the two orthogonal ports of the coil, the position of the zero electric-field plane can also be adjusted. Although implant heating is reduced with this method, a linear birdcage coil results in a whole-volume average specific absorption rate that is twice that of a quadrature birdcage coil. To solve this issue, we propose alternative methods to design implant-friendly RF coils with optimized electromagnetic fields and reduced whole-volume average specific absorption rate. With these methods, the transmit field was modified to reduce RF heating of implants and obtain uniform transmit sensitivity.
منابع مشابه
RF heating due to conductive wires during MRI depends on the phase distribution of the transmit field.
In many studies concerning wire heating during MR imaging, a "resonant wire length" that maximizes RF heating is determined. This may lead to the nonintuitive conclusion that adding more wire, so as to avoid this resonant length, will actually improve heating safety. Through a theoretical analysis using the method of moments, we show that this behavior depends on the phase distribution of the R...
متن کاملAssessment of the Characteristics of MRI Coils in Terms of RF Non-Homogeneity Using Routine Spin Echo Sequences
Introduction: One of the major causes of image non-uniformity in MRI is due to the existence of non-homogeneity in RF receive and transmit. This can be the most effective source of error in quantitative studies in MRI imaging. Part of this non-homogeneity demonstrates the characteristics of RF coil and part of it is due to the interaction of RF field with the material being imaged...
متن کاملInvestigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging
Deep Brain Stimulation (DBS) is increasingly used to treat a variety of brain diseases by sending electrical impulses to deep brain nuclei through long, electrically conductive leads. Magnetic resonance imaging (MRI) of patients pre- and post-implantation is desirable to target and position the implant, to evaluate possible side-effects and to examine DBS patients who have other health conditio...
متن کاملRadiative principles for efficient RF transmission at UHF : radiative antennas and travelling waves
RF coils can be classified in RF receive coils and RF transmit coils. RF receive coil are designed to create a maximal sensitivity to the NMR signals emitted by the spins while picking up a minimal amount of thermal noise. The function of RF transmit coils is to generate a homogeneous B1+ field in the imaging region while inducing minimal RF power deposition that results in undesired RF tissue ...
متن کاملEffect of body coil electric field distribution on receive-only surface coil heating.
Although in the design of transmit RF coils, B(1) homogeneity is crucial for good image quality, discussion of electric field (E-field) distribution in the literature has been mostly limited to specific absorption rate (SAR) and patient loading (dielectric) effects. In this work, we report on a different aspect of E-field: the receive-only surface coil heating resulting from the voltage drop ac...