Optical Calibration of a Submicrometer Magnification Standard
نویسندگان
چکیده
The calibration of a new submicrometer magnification standard for electron microscopes is described. The new standard is based on the width of a thin thermal-oxide film sandwiched between a silicon single-crystal substrate and a polysilicon capping layer. The calibration is based on an ellipsometric measurement of the oxide thickness before the polysilicon layer is deposited on the oxide. The uncertainty in the derivation of a thickness for the layer from the ellipsometric parameters is also derived.
منابع مشابه
Erratum: Optical Calibration of a Submicrometer Magnification Standard
[This corrects the article on p. 267 in vol. 97.].
متن کاملAn Accurate Surface Measurement Method Based on Depth-from-focus Considering Magnification Variations
Errors resulting from magnification variations of optical system are largely generated in threemensional shape measurements based on depth-from-focus. Focus measures are wrongly or ambiguously tracted, due to the variations. In this paper, a DFF-based methodology considering these magnification riations is proposed to make accurate measurements in surface morphology with high depth discontinuit...
متن کاملCalibration Markers for Digital Templating in Total Hip Arthroplasty
Digital templating with external calibration markers is the standard method for planning total hip arthroplasty. We determined the geometrical basis of the magnification effect, compared magnification with external and internal calibration markers, and examined the influence on magnification of the position of the calibration markers, patient weight, and body mass index (BMI). A formula was der...
متن کاملSubmicrometer resolution hard X-ray holography with the asymmetric bragg diffraction microscope.
The asymmetric Bragg diffraction microscope is a novel x-ray microscope which forms a magnified in-line near-field hologram by asymmetric reflection from two crossed flat crystals. In this paper, the optics of the microscope is studied theoretically. The optical transfer function is obtained, and the limiting spatial resolution, rated at 25% modulation transfer, is determined to be 0.30 microm ...
متن کاملB 707833 C : " In situ optical monitoring of RDX nanoparticles formation during rapid expansion of supercritical CO
Appendix A. Calibration of the sensitivity of the ICCD camera Two independent techniques were used for the sensitivity calibration: Rayleigh scattering from air and Mie scattering from standard spherical polystyrene latex particles. The laser beam was focused by a spherical lens (focal length 200 mm at 248 nm), so that the focal point was in the field of view of the optical system composed of t...