Optimization of the Scan Protocol for the Reduction of Diaphragmatic Motion Artifacts Depicted on CT Angiography: a Phantom Study Simulating Pediatric Patients with Free Breathing
نویسندگان
چکیده
OBJECTIVE This study was designed to optimize the scan protocol of CT angiography to reduce diaphragmatic motion artifacts in pediatric patients with free-breathing. MATERIALS AND METHODS A phantom with twelve tubes with different diameters was constructed. To simulate free-breathing, the phantom was connected to a motor, and the phantom moved along the axis of scan. Scans were performed under several conditions: different pitch (1, 1.5) and gantry rotation time (0.37 and 0.75 sec), and different movement range (1 cm, 3 cm) and rates (20/min, 40/min). For CT scanning, a 16-channel CT scanner was used and fixed factors of the CT protocol were as follows: 100 effective mAs, 80 kVp, reconstruction with a soft-algorithm, beam collimation 16 x 75 mm, reconstruction thickness of 1 mm, and an interval of 0.5 mm. CT scans were repeated five times. Each tube was evaluated with the use of a grading system (0 for images where tubes were not discriminable and 2 for images where tubes were clearly discriminable). RESULTS A higher pitch and shorter gantry rotation time produced images with a higher grade. Average grades for the higher pitch (1.5) and faster gantry rotation time (0.37 sec) for each combination of movement were as follows: 1.94 (range 1 cm and rate 20/min), 1.42 (range 1 cm and rate 40/min), 0.86 (range 3 cm and rate 20/min) and 0.52 (range 3 cm and rate 40/min). Average grades for the lower pitch (1) and slower gantry rotation time (0.75 sec) for each combination of movement were 1.08, 0.56, 0.32 and 0.08, respectively. CONCLUSION The scanning speed and especially the pitch are important parameters for CT scans to overcome a respiratory motion artifact.
منابع مشابه
Optimization of Radiation Dose in Cranial Computed Tomography among Adults: Assessment of Radiation Dose against Image Quality
Introduction: The rapid use of computed tomography (CT) scan is of great concern, due to increase in patients’ dose. Optimization of CT protocol is a vital issue in dose reduction. This study aimed to optimize radiation dose in cranial CT and assess modifications in image quality under radiation dose reduction. Material and Methods: A poly(me...
متن کاملحفاظت پرتویی بیماران در سی تی آنژیوگرافی قلب
Introduction: Cardiac CT (Computed Tomography) angiography applies as a pre-exam method for evaluating the patients with medical problems in their cardiac arteries. Due to the high delivered dose in this imaging method, dose reduction techniques and imaging strategies that serve the images quality with minimum patient dose are the subjects for investigation and research studies. This study will...
متن کاملDesign and manufacturing a CTDI pediatric phantom for CT dosimetry using TLD, ion chambers, and gel dosimetry
Introduction: Nowadays, the demand for CT-scan is increased, and this issue has become a critical subject in radiation protection, because of the risk of cancer especially for children. The quality assurance of CT images is a necessary factor that can effect on children dose. The CT dose index is an important dosimetry index in CT-Scan dosimetry. The purpose of this study is t...
متن کاملEvaluation of Metal Artifact Reduction software in Computed Tomography
Introduction: The image quality of computed tomography (CT) can be seriously lowered by metal implants of patients. These implants are known to exert a significant impact on diagnostic accuracy due to artifacts. The current study aimed to assess the usefulness of Metal Artifact Reduction (MAR) software in the reduction of metal artifacts, in comparison to iterative rec...
متن کاملOptimization of Parameters in 16-slice CT-scan Protocols for Reduction of the Absorbed Dose
Introduction In computed tomography (CT) technology, an optimal radiation dose can be achieved via changing radiation parameters such as mA, pitch factor, rotation time and tube voltage (kVp) for diagnostic images. Materials and Methods In this study, the brain, abdomen, and thorax scaning was performed using Toshiba 16-slice scannerand standard AAPM and CTDI phantoms. AAPM phantom was used for...
متن کامل