A mechanical study investigating the use of two different methods (grub and bolt screws) to secure external fixation half pins to circular frames. A four part experiment: (1) Grub and bolt screws were used to secure half pins in Taylor Spatial frames. Loosening torques were measured using a calibrated torque wrench. (2) Using universal testing machine (UTM), axial loading was applied to establish thresholds for loosening in grub and bolt screw constructs. (3) We established the application torque to produce failure at the head-driver interface using these two methods. (4) Grub and bolt screw constructs were created controlling torque. Using UTM, axial loading was applied to establish thresholds for loosening. Statistical analysis was conducted using SPSS v20.0.0. (1) Higher torque is employed when bolt rather than grub screws is used to secure half pins on Rancho cubes (p < 0.05). (2) Loading threshold for loosening is higher in bolt screw constructs when the torque applied to secure the constructs is not controlled (p < 0.05). (3) Torque required for failure at the head-driver interface was 5.3 Nm for grub screws and 9.9 Nm for bolts. (4) Loading threshold for loosening is higher in grub screw constructs when the same torque was applied to secure them (p < 0.05). Bolt screws can be employed to secure the half pin-frame interface. They offer good stability and reduce failure at the head-driver interface. Further research is needed to determine the mechanical properties of such constructs in vivo.
منابع مشابه
Biomechanical analysis of four external fixation pin insertion techniques
Having multiple external fixation pin designs and insertion techniques has led to debate as to which combination creates the stiffest construct. This study sought to biomechanically evaluate construct strength using self-drilling (SD) and self-tapping (ST) pins inserted with either bicortical or unicortical fixation. SD and ST 5.0 mm stainless steel pins were used in combination with bicortical...
متن کاملParametric analyses of pin-bone stresses in external fracture fixation devices.
Problems occurring in the use of external fixators for bone fractures include pin-bone interface necrosis, infection, and loosening. These may be initiated and enhanced by pin-bone interface stress levels. Based on stress data from finite element method (FEM) models, an analytical "closed-form" model of the local pin-bone configuration in long-bone fracture fixation is developed. This model is ...
متن کاملTreatment of Open Tibial Fractures: Convert-ing or Continuing External Fixation?
Background: The treatment of open tibial fractures is still an orthopaedic challenge and full of complications. In many cases the use of external fixator that has been known as a non-union machine is obligatory with a high incidence of pin track infection and other complications. The aim of this study was to compare the use of external fixation as a definite method for treatment of open tibial ...
متن کاملThe Half - Pin and the Pin Tract :
A lthough the key principles of external fixation have changed little over the years, there remains a significant amount of variation in fixation methods and postoperative care. In particular, use and management of halfpins intended for prolonged reconstruction are the subject of strong opinions and intense debate. Half-pins are commonly available in stainless steel and titanium. They are also ...
متن کاملTemporary external fixation is safe in a combat environment.
BACKGROUND External fixation has been used extensively during recent wars as a damage control measure for fractures in coalition forces being evacuated. We hypothesize that external fixation is a safe and effective initial stabilization procedure for combat-related open fractures. METHODS Records on 55 consecutive type III tibia fractures between March 2003 and September 2007 were reviewed. W...