Severe metallosis following oxidised zirconium wear in total hip arthroplasty.
نویسندگان
چکیده
To cite: Gkouliopoulou E, Agathangelidis F, Vampertzis T, et al. BMJ Case Rep Published online: [please include Day Month Year] doi:10.1136/bcr-2016218025 DESCRIPTION We present a case of a man aged 81 years who underwent a primary cementless total hip arthroplasty. Eight years after the index surgery, failure of the arthroplasty was revealed by the presence of the radiographic ‘cloud sign’ (figure 1). The original components used were the R3 acetabular cup with a Synergy femoral stem, highly crossed linked polyethylene liner and an oxidised zirconium (Oxinium) femoral head (Smith & Nephew Synergy, Memphis, Tennessee, USA). The initial investigation included a CT scan, full blood tests and a tissue biopsy which revealed no malignancy. Subsequently, during revision surgery, there was extensive osteolysis with metallosis. There was black staining of the periprosthetic soft tissues and the hip pseudocapsule was filled with thick black fluid (figure 2). There was severe wear of the femoral head and the polyethylene liner. Following debridement and removal of the components, a revision cementless stem and acetabular cup was used to reconstruct the hip joint. The cloud sign in the plain radiographs—also known as the bubble sign—is an indication of soft tissue metallosis. The findings are usually subtle and the debris is well tolerated by the patients. A few cases with characteristic radiographic findings are reported in the literature, but to the best of our knowledge, the present case is the most impressive due to the extent of the metallosis. 3 Oxidised zirconium is a metallic alloy with a ceramic surface that combines the best of metal and ceramics. However, if the outer surface of the head is damaged, then rapid destruction of the head can occur.
منابع مشابه
The Result of Revision Total Hip Arthroplasty in Patients with Metallosis Following a Catastrophic Failure of a Polyethylene Liner
BACKGROUND Wear cannot be completely prevented after total hip arthroplasty. If severe polyethylene (PE) liner wear develops, the so-called catastrophic failure occurs and metallosis develops. We postulated that longevity of the new implant may be affected after revision surgery for metallosis following a catastrophic failure of a PE liner due to the substantial amount of PE wear particles and ...
متن کاملMetallosis following full thickness wear in total hip arthroplasty
Extreme wear through the metal-backed acetabular cup following total hip arthroplasty is rare, as symptoms such as pain and disability will usually manifest earlier. We present the second reported case of extreme wear in a 78-year-old male patient, who presented 20 years following an uncemented total hip replacement with a 3-year history of hip pain, clicking on ambulation and worsening mobilit...
متن کاملSevere metallosis leading to femoral head perforation.
This article describes a case of severe metallosis in a 67-year-old woman who initially underwent primary total hip arthroplasty with a ceramic-on-ceramic articular bearing. This was subsequently revised to a metal-on-polyethylene articulation due to ceramic liner fracture. She presented with severe hip pain and a pelvic mass. Infective workup was negative. Perforation of the cobalt-chrome femo...
متن کاملFive-year comparison of wear using oxidised zirconium and cobalt-chrome femoral heads in total hip arthroplasty: a multicentre randomised controlled trial.
Oxidised zirconium (OxZi) has been developed as an alternative bearing surface for femoral heads in total hip arthroplasty (THA). This study has investigated polyethylene wear, functional outcomes and complications, comparing OxZi and cobalt-chrome (CoCr) as part of a three-arm, multicentre randomised controlled trial. Patients undergoing THA from four institutions were prospectively randomised...
متن کاملGranulomatous Lung Disease: A Novel Complication following Metallosis from Hip Arthroplasty
A case of a female patient with local and systemic complications of metallosis, following catastrophic wear of a revised hip arthroplasty, is presented. The patient had a history of a fractured ceramic-on-ceramic implant, exchanged with a metal-on-polyethylene prosthesis. Systemic complications included sarcoidosis-like reactions, presenting as granulomatous lung disease, along with chorioretin...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- BMJ case reports
دوره 2016 شماره
صفحات -
تاریخ انتشار 2016