Metallic Discoloration on the Right Shin Caused by Titanium Alloy Prostheses in a Patient with Right Total Knee Replacement
نویسندگان
چکیده
The differential diagnosis of bluish-gray pigmentation of the skin includes dermal melanocytosis, malignant melanoma and metallic discoloration. Of these disorders, metallic discoloration, which develops from deposits of fine metallic particles on the skin, is an uncommon dermatologic disease. We report a case of metallic discoloration of the right shin caused by titanium alloy prostheses in a patient with right total knee replacement. The patient had bluish-gray pigmentation on her right shin as well as pain while walking. Histopathologic findings of the skin lesion showed abundant black particles in the perivascular area of the lower dermis. Polarizing microscopy revealed refractile foreign materials in the dermis. An energy dispersive X-ray spectrophotometer unit attached to a transmission electron microscope detected titanium and aluminium peaks in the specimen.
منابع مشابه
Investigating the Impact of Knee Prosthesis in Patients’ Body on Radiation Dose Distribution: A Monte Carlo Approach
Introduction: Metal prostheses in patients affect the radiotherapy dose distribution. Metal prostheses with high density and atomic number cause major changes in scattering and attenuation of radiation. The present study aims to assess the impact of metal knee prosthesis with various dimensions and materials on radiotherapy dose distribution. Material and Methods: In this research, the Varian L...
متن کاملMetallosis of the Resurfaced Hip
Metallosis of the hip is usually defined as aseptic fibrosis, local necrosis, or loosening of the prosthesis secondary to metallic corrosion and release of wear debris.8 It has also been defined as gray discoloration of the tissues of the joint, pain, an effusion, and elevated serum metal levels. Metallosis has been found with stainless steel, titanium, and cobalt chromium alloy femoral prosthe...
متن کاملDiscoloration of titanium alloy in acidic saline solutions with peroxide.
The objective of this study was to compare corrosion behavior in several titanium alloys with immersion in acidulated saline solutions containing hydrogen peroxide. Seven types of titanium alloy were immersed in saline solutions with varying levels of pH and hydrogen peroxide content, and resulting differences in color and release of metallic elements determined in each alloy. Some alloys were ...
متن کاملBrucella Arthritis Following Total Knee Arthroplasty in a Patient with Hemophilia: A Case Report
Total knee arthroplasty (TKA) is a rewarding procedure in patients with hemophilia and end stage knee hemophilicarthropathy. However, this procedure might be associated with complications such as infection. There periprostheticjoint infection in patients with hemophilia is very well known, though we are not aware of any previous report on Brucellainfection in this group of patients. Here, we re...
متن کاملRight ventricular Hemodynamic Alteration after Pulmonary Valve Replacement in Children with Congenital Heart Disease
Introduction: In patients who underwent surgery to repair Tetralogy of Fallot, right ventricular dilation from pulmonary regurgitation may be result in right ventricular failure, arrhythmias and cardiac arrest. Hence, pulmonary valve replacement may be necessary to reduce right ventricular volume overload. The aim of present study was to assess the effects of pulmonary valve replacement on rig...
متن کامل