Point of Care D-Dimer Testing in the Emergency Department: A Bioequivalence Study
نویسندگان
چکیده
BACKGROUND D-dimer is used widely as a diagnostic aid in low- and moderate-risk patients with suspected venous thromboembolism (VTE). While our laboratory utilizes VIDAS D-dimer analyzer (bioMérieux SA, France), our emergency department (ED) recently procured a D-dimer analyzer AQT90 FLEX (Radiometer Medical ApS, Denmark) for point of care testing (POCT) to facilitate patient management. We aimed to determine whether the time taken to receive D-dimer results using the 2 different analyzers differed significantly and to quantify the limits of agreement between the results of the 2 methods measured on the same patient. METHODS Adult patients presenting to the ED and requiring diagnostic workup for suspected VTE were included in this prospective observational study. Patients underwent simultaneous D-dimer measurements using the 2 different analyzers. RESULTS The paired results from 104 patients were analyzed. The median time for the D-dimer results from triage by VIDAS was 258 min (Inter-quartile range [IQR], 173-360) and by POCT was 146 min (IQR, 55-280.5); the median time difference was 101.5 min (IQR, 82-125.5). On an average, POCT D-dimer values were 15% lower on the same sample (limits of agreement, 34-213%). POCT predicted 83% of VIDAS positive results (sensitivity, 83.3% [95% confidence interval (CI), 70.4-91.3%]; specificity, 100% [95% CI, 93.6-100%]). All patients with positive imaging were identified correctly by both methods. CONCLUSIONS POCT delivers D-dimer results in significantly shorter turnaround times than pathology services; however, poor bioequivalence between VIDAS and POCT raises the issue of acceptability for use in the ED.
منابع مشابه
Response to the Comments on 'Point of Care D-Dimer Testing in the Emergency Department-A Bioequivalence Study' and Erratum to the Results
We have read with interest the comments made on our study titled ‘Point of care (POC) D-dimer testing in the Emergency department–a bioequivalence study’ [1]. Ekelund and Heilmann, referring to the results and conclusions of the study, have suggested that one of the discrepancies may be caused by a typographical error, and this inference probably stems from the use of the term ‘range’ in the pu...
متن کاملEvaluation of Correlation between D-dimer Levels with Disease Severity in Patients with Covid-19 Referred to Hamedan Besat Hospital in 2021-2022
Introduction: The Covid-19 virus affects the hemostasis system in infected people by causing thrombosis and it increases the level of D_dimer. This study was designed and conducted for the evaluation of correlation between D-dimer levels with disease severity in patients with Covid-19 referred to Hamedan Besat hospital. Methods and Materials: 366 patients with Covid-19 referred to Besat Hospit...
متن کاملComments on Point of Care D-Dimer Testing in the Emergency Department: A Bioequivalence Study
Dear Editor We read with interest the article “Point of Care D-Dimer Testing in the Emergency Department: A Bioequivalence Study” by Perveen et al. published in Ann Lab Med 2013;33:34-38 [1]. In this article, the authors compared the performance of the D-dimer assays on the AQT90 FLEX analyzer (Radiometer Medical Aps, Åkandevej, Denmark) and the VIDAS analyzer (bioMérieux SA, RCS Lyon, France)....
متن کاملPOCT is a true asset in the emergency department
Background To determine patient priority and degree of urgency with an objective high-quality evaluation, general Point of Care testing (POCT) was established as a novel facility in the Emergency department at Holbæk Hospital. The hardware consists of AQT-90 Flex (CRP, D-dimer), ABL-800 (Na, K, Hgb, BG etc.) and HemoCue (Leukocytes), thus rendering useful parameters within initial assessment. A...
متن کاملPerspectives on cost and outcomes for point-of-care testing.
Point-of-care testing (POCT) is usually more expensive on a unit-cost basis than testing performed in a central laboratory. It is difficult to manage POCT and to maintain regulatory compliance, especially in large institutions. However, some POCT technologies have improved patient outcomes (patient self-glucose monitoring in the home, tight glycemic control in intensive care settings) or hospit...