A four-year experience with patient individualized heparin and protamine dosing using the Hemochron RxDx system.
نویسندگان
چکیده
Cardiac surgical case histories, collected over 4 years at Huntsville Hospital in Alabama, were reviewed for 2,293 patients. Patients were separated into two dosing groups for both heparin and protamine, hospital empirically dosed and Hemochron RxDx dosed. Review of the heparin dosing information found that incomplete data were collected for 47 patients, leaving 2,246 patients eligible to be evaluated for heparin dose comparison. Both RxDx recommended and empirically calculated doses were recorded, as well as the actual dose given. Of the 2,246 patients, 1671 were administered heparin according to the RxDx calculated dose, and the remaining 575 patients were dosed according to the hospital's empirical protocol. The average RxDx calculated heparin dose was 17% greater then the empirically calculated heparin dose (350 U/kg) (p < .001). Anticoagulation to target ACT (480 sec) was achieved in 92% of the patients dosed according to the RxDx recommended dose; whereas, in the empirically dosed patient group only 80% of the patients reached the target ACT after initial heparin bolus dose. Incomplete protamine dosing data was recorded for 336 patients, leaving a total of 1,957 patients available for protamine dose evaluation. All patients had an RxDx protamine calculation, empirical protamine calculation, and actual amount of protamine dosed recorded. Of the 1,953 patients, 1,764 were dosed according to the RxDx recommended dose, with the remaining 189 patients dosed empirically (1 mg protamine/100 U of heparin). In both the RxDx and the empirical groups, 96% of the patients returned to baseline following initial protamine infusion. The overall RxDx dose (293 mg) was 16% lower than the average empirical dose (348 mg). The RxDx system has been shown to be an effective method for determining patient-specific dosing for both heparin and protamine. This long-term clinical experience demonstrates the consistency and reliability of patient maintenance using this individualized dosing system, which has been shown, in other independent evaluations, to lead to improved patient outcomes.
منابع مشابه
The Hemochron Response RxDx heparin and protamine dosing system.
The use of dosing assays to calculate heparin and protamine dose requirements during cardiac surgery has been shown to significantly improve overall postoperative patient outcome. When patients are managed with an individualized dosing system, intraoperative and postoperative transfusion requirements and bleeding are reduced. The Hemochron RxDx system is widely used as a complement to tradition...
متن کاملHand-Held Personal Digital Assistant Program for the HEMOCHRONregistered RxDxregistered Heparin and Protamine Dosing System
The use of in vitro dosing assays for heparin and protamine during cardiac surgery has significantly improved overall postoperative patient outcome. The HEMOCHRON RxDx system (International Technidyne Corp, Edison, NJ) is widely used for anticoagulation management. Based on a series of consecutive in vitro tests, the RxDx system is used to quantify the patient’s heparin requirement (heparin res...
متن کاملIncreased accuracy in heparin and protamine administration decreases bleeding: a pilot study.
Three to 5 percent of the patients undergoing cardiac surgery are reoperated because of bleeding. When a surgical cause can be excluded, heparin/protamine mismatch may be considered. Insufficient reversal of heparin and overdosing of protamine may cause postoperative bleeding. The purpose of the study was to evaluate whether a heparin-protamine titration system, Hemochron RxDx, could reduce pos...
متن کاملUtility of In Vitro Heparin and Protamine Titration for Dosing During Cardiopulmonary Bypass Surgery
Methods used to maintain a hemostatic balance during cardiopulmonary bypass surgery include the optimization of heparin and protamine dosing. Higher heparin doses have been associated with increased bleeding, while other studies have attributed reduced blood loss and transfusion requirements to lower protamine doses. We have evaluated the use of an in vitro heparin and protamine titration syste...
متن کاملComparison of Celite and Kaolin Based Heparin and Protamine Dosing Assays During Cardiac Surgery: The In Vitro Effect of Aprotinin
Both aprotinin and in vitro heparin and protamine dosing assays (RxDx System) have been shown to reduce postoperative bleeding in the cardiac surgery patient. While celite activated clotting time tests in the presence of heparin are artificially prolonged by aprotinin, the kaolin based ACT is not. In order to optimize heparin anticoagulation management during cardiac surgery when aprotinin is u...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The journal of extra-corporeal technology
دوره 33 1 شماره
صفحات -
تاریخ انتشار 2001