Failure of Polyethelene Insert Locking Mechanism after a Posterior Stabilised Total Knee Arthroplasty- A Case Report
نویسندگان
چکیده
INTRODUCTION Disengagement of polyethylene insert used in total knee arthroplasty is a rare but serious complication. Still rarer is disengagement because of failure of tibial insert locking mechanism. We report a previously unpublished complication of polyethylene insert locking mechanism failure in a 10-months-old posterior stabilized total knee arthroplasty in a 70-year-old woman with osteoarthritis for whom Attune (Depuy) knee implant was used. CASE PRESENTATION A 70-year-old female underwent (Attune, Depuy) primary bilateral posterior stabilised total knee arthroplasty in a private hospital. The patient did not have any complaints and had had been functioning well post her arthroplasty. After five months of surgery she had a fall and sustained injury over right hip which was treated with Cemented Bipolar Hemiarthroplasty. Ten months after index surgery, she sustained trivial fall and presented to the same hospital with knee pain and swelling, where the right knee prosthesis was found to be dislocated. An attempted closed reduction under anaesthesia failed, after which she was referred to our centre with an unstable, painful, swollen right knee in a long knee brace. The physical examination at the time of admission showed posterior sag of the tibia, fullness in the postero-lateral corner, quadriceps muscle atrophy without any neurovascular deficit oflower leg. Postero-lateral dislocation was confirmed with radiographs. Surgical error as a possible causative factor was excluded because patient had been functioning well after surgery. Her comorbidities included hypertension and hyponatremia. ESR and CRP were within normal limits. An open reduction surgery was planned. On exposure, polyethylene was found in the postero-lateral corner of the knee. We were not sure that revising the polyethylene alone would suffice as the poly and locking mechanism was of a relatively new design and hence it was decided to proceed with revision of the components. Revision was done with stemmed components, distal femoral augments and a constrained prosthesis (Total Condylar 3, Depuy). Intraoperative cultures were negative. The patient had an uncomplicated post-operative course. CONCLUSION Disengagement of polyethylene should be considered as a differential diagnosis in patients who present with acute swelling and instability of the knee. Though it's a rare complication, there is a possibility that design of the implant or its locking mechanism could contribute to dislocations in future.
منابع مشابه
Nontraumatic, spontaneous dislocation of polyethylene tibial insert 1 year after TKA.
The authors report a case of nontraumatic, spontaneous dislocation of a polyethylene insert detected 1 year after total knee arthroplasty. The patient demonstrated initial improvement and returned to work 4 months postoperatively. At 6 months postoperatively, the patient developed pain and a clunking sensation with motion; however, he denied any traumatic precipitating events. An arthroscopic p...
متن کاملPost impingement in posterior stabilised total knee arthroplasty.
Posterior stabilised implants are a well-proven treatment for patients requiring primary total knee arthroplasty. Concerns about the posterior stabilised (PS) design have been raised and recent studies suggest that the post-cam articulation can be an additional source of polyethylene wear debris. In this study, the authors report impingement of the tibial post against the patellar component in ...
متن کاملFracture of the Anterior Locking Flange of a Total Knee Arthroplasty Polyethylene Liner Presenting with Pain following Knee Replacement
INTRODUCTION Fracture of the modern polyethylene insert of a total knee arthroplasty is rare. We describe the first case of a fractured anterior locking flange of the commonly used Depuy Press-fit Condylar (PFC) Sigma prosthesis. CASE REPORT The 80 year old Caucasian gentleman presented 8 years following previously uncomplicated and successful primary total knee replacement with pain, swellin...
متن کاملIn vivo kinematics of total knee arthroplasty. Concave versus posterior-stabilised tibial joint surface.
We studied the kinetics of the knee in 20 patients (22 knees) 12 months after total knee arthroplasty (TKA), by using three-dimensional radiostereometry and film-exchanger techniques. Eleven knees had a concave (constrained) tibial implant and 11 a posterior-stabilised prosthesis. Eleven normal knees served as a control group. In the posterior-stabilised knees there was less proximal and poster...
متن کاملTotal Knee Arthroplasty in Severe Unstable Knee: Case-Report and Literature Review
Multiplanar or global laxity in arthritic knee is rare , most of this patients have neuromuscular disorder (post poliomyelitis , spinal dystrophy) or history of knee trauma. Ligament insufficiency and severe bone loss is significant in this patient. The estimated prevalence for the concurrence of charcot marie-tooth (CMT) with myasthenia gravis (MG) suggests an extremely rare event. We have p...