Rounded leaf end modeling in Pinnacle VMAT treatment planning for fixed jaw linacs

نویسندگان

  • Lori A. Young
  • Fei Yang
  • Ning Cao
  • Juergen Meyer
چکیده

During volume-modulated arc therapies (VMAT), dosimetric errors are introduced by multiple open dynamic leaf gaps that are present in fixed diaphragm linear accelerators. The purpose of this work was to develop a methodology for adjusting the rounded leaf end modeling parameters to improve out-of-field dose agreement in SmartArc VMAT treatment plans delivered by fixed jaw linacs where leaf gap dose is not negligible. Leaf gap doses were measured for an Elekta beam modulator linac with 0.4 cm micro-multileaf collimators (MLC) using an A16 micro-ionization chamber, a MatriXX ion chamber detector array, and Kodak EDR2 film dosimetry in a solid water phantom. The MLC offset and rounded end tip radius were adjusted in the Pinnacle treatment planning system (TPS) to iteratively arrive at the optimal configuration for 6 MV and 10 MV photon energies. Improvements in gamma index with a 3%/3 mm acceptance criteria and an inclusion threshold of 5% of maximum dose were measured, analyzed, and validated using an ArcCHECK diode detector array for field sizes ranging from 1.6 to 14 cm square field arcs and Task Group (TG) 119 VMAT test cases. The best results were achieved for a rounded leaf tip radius of 13 cm with a 0.1 cm MLC offset. With the optimized MLC model, measured gamma indices ranged between 99.9% and 91.7% for square field arcs with sizes between 3.6 cm and 1.6 cm, with a maximum improvement of 42.7% for the 1.6 cm square field size. Gamma indices improved up to 2.8% in TG-119 VMAT treatment plans. Imaging and Radiation Oncology Core (IROC) credentialing of a VMAT plan with the head and neck phantom passed with a gamma index of 100%. Fine-tune adjustments to MLC rounded leaf ends may improve patient-specific QA pass rates and provide more accurate predictions of dose deposition to avoidance structures.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Determining the optimal dosimetric leaf gap setting for rounded leaf‐end multileaf collimator systems by simple test fields

Individual QA for IMRT/VMAT plans is required by protocols. Sometimes plans cannot pass the institute's QA criteria. For the Eclipse treatment planning system (TPS) with rounded leaf-end multileaf collimator (MLC), one practical way to improve the agreement of planned and delivered doses is to tune the value of dosimetric leaf gap (DLG) in the TPS from the measured DLG. We propose that this ste...

متن کامل

Development of an optimum photon beam model for head‐and‐neck intensity‐modulated radiotherapy

Intensity-modulated radiotherapy (IMRT) for complex sites such as tumors of the head and neck requires a level of accuracy in dose calculation beyond that currently used for conformal treatment planning. Recent advances in treatment planning systems have aimed to improve the dose calculation accuracy by improving the modeling of machine characteristics such as interleaf leakage, tongue and groo...

متن کامل

On using the dosimetric leaf gap to model the rounded leaf ends in VMAT/RapidArc plans

Partial transmission through rounded leaf ends of Varian multileaf collimators (MLC) is accounted for with a parameter called the dosimetric leaf gap (DLG). Verification of the value of the DLG is needed when the dose delivery is accompanied by gantry rotation in VMAT plans. We compared the doses measured with GAFCHROMIC film and an ionization chamber to treatment planning system (TPS) calculat...

متن کامل

Validation of physics improvements for IMRT with a commercial treatment‐planning system

A new Pinnacle 3D treatment-planning system software release has recently become available (v7.4, Philips Radiation Oncology Systems, Milpitas, CA), which supports modeling of rounded multileaf collimator (MLC) leaf ends; it also includes a number of other software enhancements intended to improve the overall dose calculation accuracy. In this report, we provide a general discussion of the dose...

متن کامل

SU-E-T-549: On Importance of Accurate Treatment Couch Modeling in SmartArc VMAT Plans.

PURPOSE We investigate the impact of a treatment couch and its modeling in a treatment planning system on accurate dose delivery in Volumetric Modulated Arc Therapy (VMAT). METHODS A cone-beam CT of a Varian couch on a Varian Clinac 21EX unit, consisting of a carbon-fiber imaging couch top and supporting rails, was acquired, and a couch model was created in Philips Pinnacle 9.0 TPS based on t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2016