Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates – An In Vitro Study

نویسندگان

  • Julia C. Hartung
  • Silke Wilitzki
  • Marta Thio-Lluch
  • Arjan B. te Pas
  • Gerd Schmalisch
  • Charles C. Roehr
  • Edgardo Szyld
چکیده

INTRODUCTION International resuscitation guidelines suggest to use positive end-expiratory pressure (PEEP) during manual ventilation of neonates. Aim of our study was to test the reliability of self-inflating bags (SIB) with single-use PEEP valves regarding PEEP delivery and the effect of different peak inflation pressures (PIP) and ventilation rates (VR) on the delivered PEEP. METHODS Ten new single-use PEEP valves from 5 manufacturers were tested by ventilating an intubated 1 kg neonatal manikin containing a lung model with a SIB that was actuated by an electromechanical plunger device. Standard settings: PIP 20 cmH2O, VR 60/min, flow 8 L/min. PEEP settings of 5 and 10 cmH2O were studied. A second test was conducted with settings of PIP 40 cmH2O and VR 40/min. The delivered PEEP was measured by a respiratory function monitor (CO2SMO+). RESULTS Valves from one manufacturer delivered no relevant PEEP and were excluded. The remaining valves showed a continuous decay of the delivered pressure during expiration. The median (25th and 75th percentile) delivered PEEP with standard settings was 3.4(2.7-3.8) cmH2O when set to 5 cmH2O and 6.1(4.9-7.1) cmH2O when set to 10 cmH2O. Increasing the PIP from 20 to 40 cmH2O led to a median (25th and 75th percentile) decrease in PEEP to 2.3(1.8-2.7) cmH2O and 4.3(3.2-4.8) cmH2O; changing VR from 60 to 40/min led to a PEEP decrease to 2.8(2.1-3.3) cmH2O and 5.0(3.5-6.2) cmH2O for both PEEP settings. CONCLUSION Single-use PEEP valves do not reliably deliver the set PEEP. PIP and VR have an effect on the delivered PEEP. Operators should be aware of these limitations when manually ventilating neonates.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Innovating Neonatal Resuscitation Bags for Ease of Use in Diverse Locations.

The Chinese emperor Huang-Ti (2698–2599 BCE) noted that newborn death from ventilatory failure was common in premature infants.1 Galen (129–199 CE) inflated the lungs of deceased animals with a bellows and concluded that the air movement causes the chest to rise.1,2 Between 1850 and 1950, many techniques were used to resuscitate newborns, including swinging the infant upside down,1-3 rhythmic t...

متن کامل

T-piece resuscitator versus self-inflating bag for preterm resuscitation: an institutional experience.

BACKGROUND Manual ventilation in the delivery room is provided with devices such as self-inflating bags (SIBs), flow-inflating bags, and T-piece resuscitators. OBJECTIVE To compare the effect of type of manual ventilation device on overall response to resuscitation among preterm neonates born at < 35 weeks gestation. METHODS Retrospective data were collected in 2 time periods. Primary outco...

متن کامل

Comparison of three manual ventilation devices using an intubated mannequin.

OBJECTIVE To compare three devices for manual neonatal ventilation. DESIGN Participants performed a two minute period of ventilation using a self inflating device, an anaesthesia bag with attached manometer, and a Neopuff device. An intubated neonatal mannequin, approximating a 1 kg infant with functional lungs, was used for the study. Target ventilation variables included a rate of 40 breath...

متن کامل

Neonatal and pediatric manual hyperinflation: influence of oxygen flow on ventilation parameters.

BACKGROUND Although self-inflating bags are widely used for manual hyperinflation, they do not allow ventilation parameters, such as pressure or volume, to be set. We studied the ventilation performance of neonatal and pediatric self-inflating bags. METHODS We asked 22 physiotherapists to manually hyperinflate 2 lung models (neonatal and pediatric), using self-inflating bags from 3 manufactur...

متن کامل

The self-inflating resuscitation bag delivers high oxygen concentrations when used without a reservoir: implications for neonatal resuscitation.

OBJECTIVE To measure the delivered fractional oxygen concentration (F(DO(2))) from preterm-size Laerdal silicone resuscitators (PLSR) without a reservoir. BACKGROUND The North American Neonatal Resuscitation Program manual states that self-inflating bags without a reservoir deliver approximately 40% oxygen, differing from the PLSR manufacturer's specifications. METHODS A neonatal test lung ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2016