Energy Dependence of the New Gafchromic EBT3 Film:Dose Response Curves for 50 KV, 6 and 15 MV X-Ray Beams

نویسندگان

  • Guerda Massillon-JL
  • Sou-Tung Chiu-Tsao
  • Ivan Domingo-Muñoz
  • Maria F. Chan
چکیده

Purpose: To study and compare the dose response curves of the new GafChromic EBT3 film for megavoltage and kilovoltage x-ray beams, with different spatial resolutions. Methods: EBT3 films (lot#A101711-02) were exposed to each x-ray beam (6 MV, 15 MV, and 50 kV) at 7 dose values (50 3200 cGy). Each film piece was scanned three consecutive times in the center of Epson 10000XL flatbed scanner in 48-bit color at two separate spatial resolutions of 75 and 300 dpi. The data were analyzed using ImageJ and, for each scanned image, a region of interest (ROI) of 2 × 2 cm at the field center was selected to obtain the mean pixel value with its standard deviation in the ROI. For each energy, dose value and spatial resolution, the average net optical density (netOD) and its associated uncertainty were determined. The Student’s t-test was performed to evaluate the statistical differences between the net OD/dose values of the three energy modalities, with different color channels and spatial resolutions. Results and Discussion: The dose response curves for the three energy modalities were compared in three color channels. Weak energy dependence was found. For doses above 100 cGy, no statistical differences were observed between 6 and 15 MV beams, regardless of spatial resolution and color channel. However, statistical differences were observed between 50 kV and the megavoltage beams. The degree of energy dependence (from MV to 50 kV) was found to be a function of color channel, dose level, and spatial resolution. Conclusions: The dose response curves for GafChromic EBT3 films were found to be weakly dependent on the energy of the photon beams from 6 MV to 15 MV. For very low energy photon (e.g. 50 kV), variation of more than 11% due to the energy-dependence is observed, depending on the absorbed dose, spatial resolution and color channel used.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams.

PURPOSE This work investigates the dose-response curves of GAFCHROMIC(®) EBT, EBT2, and EBT3 radiochromic films using synchrotron-produced monochromatic x-ray beams. EBT2 film is being utilized for dose verification in photoactivated Auger electron therapy at the Louisiana State University Center for Advanced Microstructures and Devices (CAMD) synchrotron facility. METHODS Monochromatic beams...

متن کامل

Gafchromic EBT3 film dosimetry in electron beams — energy dependence and improved film read‐out

For megavoltage photon radiation, the fundamental dosimetry characteristics of Gafchromic EBT3 film were determined in 60Co gamma ray beam with addition of experimental and Monte Carlo (MC)-simulated energy dependence of the film for 6 MV photon beam and 6 MeV, 9 MeV, 12 MeV, and 16 MeV electron beams in water phantom. For the film read-out, two phase correction of scanner sensitivity was appli...

متن کامل

Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range

The objective of this study is to evaluate the suitability of recently introduced radiochromic film EBT3 for clinical dosimetry in the kilovoltage (kV) range. For this purpose, a kV X-ray irradiator, X RAD 320ix in the range 70 to 300 kVp, a clinical 60Co source, a 6 MV and an 18 MV X-ray clinical beam from a Varian linear accelerator were calibrated following AAPM dosimetry protocols. EBT3 fil...

متن کامل

SU-E-T-142: Evaluation of the Energy Dependence of Gafchromic EBT3 Film for Electron Beams.

PURPOSE To evaluate the energy dependence of Gafchromic EBT3 film for establishing a quality assurance method of bolus electron conformal radiotherapy. METHODS We irradiated electron beam to EBT3 films, which were set in the water tank. The linear accelerator used was Varian Clinac 21EX. The energy of electron beams were 9 and 12 MeV. The irradiated field size was 10×10 cm2 and the source to ...

متن کامل

Investigation of surface and buildup region doses for 6 MV high energy photon beams in the presence of a thermoplastic mask

Background: The accuracy and the reproducibility of radiotherapy can be provided by using immobilization devices such as thermoplastic masks. In head and neck cancer radiotherapy, the patients are mostly immobilized by using thermoplastic masks. In this study, the effect of the thermoplastic mask to the surface and buildup region doses was investigated by using Markus parallel plate ion chamber...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2013