Intracavitary vaginal brachytherapy using a custom balloon applicator
نویسندگان
چکیده
A custom balloon applicator was created to deliver intracavitary high-dose-rate brachytherapy to a patient with a superficial vaginal cancer. The patient was unable to be treated conventionally due to an extremely narrow introitus that prevented the introduction of a conventional cylinder applicator. The custom applicator was constructed by inserting a straight titanium tandem applicator through the drainage lumen of a Foley catheter. The applicator was inserted and the catheter balloon was inflated when positioned at the vaginal apex. Three brachytherapy treatments were performed using this technique. Individual balloon eccentricities resulted in small radial tandem offsets within the balloon. This phenomenon was exploited by orientating the tandem offset in the direction of the target volume. Acceptable dosimetry was achieved for all fractions and the procedure was very well tolerated. The custom applicator was a viable solution that was safely developed in a short time frame using equipment readily available in our department.
منابع مشابه
Clinical applications of custom-made vaginal cylinders constructed using three-dimensional printing technology
PURPOSE Three-dimensional (3D) printing technology allows physicians to rapidly create customized devices for patients. We report our initial clinical experience using this technology to create custom applicators for vaginal brachytherapy. MATERIAL AND METHODS Three brachytherapy patients with unique clinical needs were identified as likely to benefit from a customized vaginal applicator. Pat...
متن کاملClinical assessment of 252Californium neutron intracavitary brachytherapy using a two-channel Y applicator combined with external beam radiotherapy for endometrial cancer
OBJECTIVE The aim of this study was to determine the efficacy of 252Californium neutron intracavitary brachytherapy using a two-channel Y applicator combined with external beam radiotherapy for the treatment of endometrial cancer. METHODS Thirty-one patients with stage I-III endometrial cancer were recruited for this study. The stage I patients received only 252Californium neutron intracavita...
متن کاملThe use of trans-applicator intracavitary ultrasonography in brachytherapy for cervical cancer: phantom study of a novel approach to 3D image-guided brachytherapy
PURPOSE To assess the feasibility of applying trans-applicator intracavitary ultrasonography to image-guided brachytherapy for cervical cancer. MATERIAL AND METHODS For this experiment, a phantom was created and included a polyethylene tube, intended to simulate a tandem applicator, which was inserted into chicken meat and embedded in agar, along with magnetic resonance imaging (MRI)-compatib...
متن کاملInterface dosimetry for electronic brachytherapy intracavitary breast balloon applicators
In this study, we evaluate the attenuation of the dose due to barium-impregnation in the region between the surface of an electronic brachytherapy (EBT) balloon applicator for accelerated partial breast irradiation (APBI) and the prescription point at 1 cm depth in tissue. To perform the study, depth dose curves were calculated using a general purpose multi-particle transport code (FLUKA) for a...
متن کاملA novel intracavitary applicator design for the treatment of deep vaginal fornices: preliminary dose metrics and geometric analysis
PURPOSE To investigate the dose distributions associated with a novel balloon sleeve placed over a standard cylinder applicator. MATERIAL AND METHODS A computed tomography (CT) scan of a sleeve balloon shaped to inflate into the vaginal fornices was used to digitize 1-, 3-, and 5-catheter configurations. Point doses for rectum, apex, and fornix were calculated and compared to the values assoc...
متن کامل