Realistic simulations on the dark rim artifact for myocardial perfusion protocols
نویسندگان
چکیده
Background and Aims Motion and Gibbs artifacts have been previously shown separately as probable sources of Dark Rim Artifacts (DRA) in myocardial perfusion imaging [1,2]. Their relative importance through the cardiac cycle for a range of typical perfusion protocols has not been fully examined. Therefore, the appearance of Gibbs, motion, and T1 and T2* k-space modulation were studied by numerically simulating typical perfusion protocols. Methods The raw-data of a short axis image of the heart was simulated in MATLAB; some data was simulated with a true perfusion defect all around the subendocardium with a third of the myocardial wall thickness. Due to the fact that the raw-data is finite and discretely sampled, all magnitude images are corrupted with Gibbs artifacts. Myocardial motion was also simulated by considering a radial motion of the endocardium during the raw-data acquisition, according to measured radial myocardial velocities by Jung et al. [3]. Three different sequences were simulated: GRE (gradient recalled echo), bSSFP (balanced steady state free precession), and hEPI (hybrid echo planar imaging) with an EPI factor of 4. The sequence protocols were (GRE/bSSFP/hEPI): TR 2.4/2.2/5.8 ms; TE 1.2/1.0/1.22 ms; Time from saturation pulse to centre of k-space 80/91/108 ms; flip angle 12/50/30 (degrees), FOV 370x370 mm, TSENSE R=2. The GRE and bSSFP used linear coverage, while the hEPI sequence k-space trajectory used is described by Ding et al [4]. The signal intensity of both the left ventricle and myocardium were taken from the first pass of contrast agent in real perfusion studies done with each sequence. Matrix sizes of 128, 160 and 192 at a range of phase-encode resolutions were simulated. All the magnitude images were zero filled to a matrix size of 2048x2048, before FFT, in order to reduce the dependency of the artifacts’ edge sub-pixel position. Motion simulations were done for central rawdata at three different times in the heart cycle: 100/300/400 ms after R-wave. This time corresponds to central rawdata coverage for the three sequences coinciding with peak-systole/end-systole/peak-diastole respectively. T1 (30/730 ms for blood/myocardium) and T2* (15/15 ms for blood/myocardium) relaxation in the raw-data were also simulated for each sequence. Some of the simulated raw-data was filtered with a Hamming filter in order to study its effects in Gibbs artifacts and real perfusion defects. Results and Discussion Figure 1 shows on the left a simulated short axis image with no motion, showing a thin DRA all around the subendocardium, Figure a. After being Hamming filtered the Gibbs artifacts are nulled, Figure b. Figure c and d are identical to a and b but with an added true defect around the subendocardium. The Hamming filter blurs slightly the true defect, although, because a high resolution of 192x192 was used, the defect is still clearly visible. On the centre it is shown a comparison for the motion artifacts simulation between a bSSFP sequence and an hEPI sequence. The results obtained for the GRE sequence (not shown) are similar to the ones obtained with the bSSFP sequence. It is shown the magnitude images for the bSSFP sequence at two different resolutions, 128x128, and 192x192, and for three different trigger times, 100/300/400 ms Figures e-j; the same results are shown, for the hEPI sequence, Figure k-p. The raw-data was not filtered; therefore Gibbs artifacts are visible in all images with the lower resolution of 128x128. As resolution increases, data acquisition time increases, thus motion artifacts become more prominent. Motion blurs the endocardial border, reducing its sharpness which reduces the visibility of Gibbs artifacts, this happens mainly for trigger times 100 ms and 400 ms. For the 300 ms trigger time, the radial velocities of the endocardium are much smaller when compared to the other two trigger times making motion artifacts less prominent and Gibbs artifacts more visible. Motion artifacts also create DRAs all around the subendocardial border except along the frequency encode direction for the bSSFP sequence at 100 and 400 ms, while the hEPI sequence shows ghosting of the bright blood pool in the phase encode direction instead of a dark rim in the endocardial region for the same trigger times. Figure q and r show hEPI magnitude images simulated with myocardial motion during diastole with the true defect at 128x128 and 192x192 matrix sizes respectively. The motion artifacts of the hEPI sequence do not render the added true defect invisible. T1 and T2* k-space modulation introduces some blurring with a spread of the point spread function of up to 25% FWHM, but no DRA is created.
منابع مشابه
Simulation of Motion-Induced Dark-Rim Artifacts for Cartesian and Spiral Pulse Sequences
Introduction: First-pass MR perfusion images may be corrupted by dark-rim artifacts, which are at least in part a consequence of myocardial motion during data acquisition [1]. We have previously shown that the non-linear time-varying displacements of the myocardium resulting from cardiac contraction result in dark-rim artifacts that have a different appearance in images obtained at different ph...
متن کاملTowards elimination of the dark-rim artifact in first-pass myocardial perfusion MRI: removing Gibbs ringing effects using optimized radial imaging.
PURPOSE Subendocardial dark-rim artifacts (DRAs) remain a major concern in first-pass perfusion (FPP) myocardial MRI and may lower the diagnostic accuracy for detection of ischemia. A major source of DRAs is the "Gibbs ringing" effect. We propose an optimized radial acquisition strategy aimed at eliminating ringing-induced DRAs in FPP. THEORY AND METHODS By studying the underlying point sprea...
متن کاملProjection imaging of myocardial perfusion: minimizing the subendocardial dark-rim artifact
Background Current methods for clinical myocardial perfusion (MP) imaging suffer from image artifacts, specifically the socalled subendocardial dark-rim artifact (DRA) [1-3]. DRAs are especially limiting since they can reduce the sensitivity/specificity of detecting subendocardial MP deficits; hence, eliminating such artifacts remains an active area of research [1]. In this work, we demonstrate...
متن کاملVariability of myocardial perfusion dark rim Gibbs artifacts due to sub-pixel shifts
BACKGROUND Gibbs ringing has been shown as a possible source of dark rim artifacts in myocardial perfusion studies. This type of artifact is usually described as transient, lasting a few heart beats, and localised in random segments of the myocardial wall. Dark rim artifacts are known to be unpredictably variable. This article aims to illustrate that a sub-pixel shift, i.e. a small displacement...
متن کاملThe truncation artifact in a skinny patient following myocardial perfusion SPECT
Myocardial perfusion SPECT is one of the most common imaging techniques performed in nuclear medicine departments. To avoid misleading interpretation, it is necessary to address the quality control and technical problems. The truncation artifact occurs when the patient size is large relative to the field of view of the camera, causing false perfusion defects in the LV myocardium, misinterpreted...
متن کاملEffect of Improving Spatial or Temporal Resolution on Image Quality and Quantitative Perfusion Assessment with k-t SENSE Acceleration in First-Pass CMR Myocardial Perfusion Imaging
k-t Sensitivity-encoded (k-t SENSE) acceleration has been used to improve spatial resolution, temporal resolution, and slice coverage in first-pass cardiac magnetic resonance myocardial perfusion imaging. This study compares the effect of investing the speed-up afforded by k-t SENSE acceleration in spatial or temporal resolution. Ten healthy volunteers underwent adenosine stress myocardial perf...