Risks of spontaneous injury and extraction of an active fixation pacemaker lead: report of the Accufix Multicenter Clinical Study and Worldwide Registry.
نویسندگان
چکیده
BACKGROUND The Telectronics Accufix pacing leads were recalled in November 1994 after 2 deaths and 2 nonfatal injuries were reported. This multicenter clinical study (MCS) of patients with Accufix leads was designed to determine the rate of spontaneous injury related to the J retention wire and results of lead extraction. METHODS AND RESULTS The MCS included 2589 patients with Accufix atrial pacing leads that were implanted at or who were followed up at 12 medical centers. Patients underwent cinefluoroscopic imaging of their lead every 6 months. The risk of J retention wire fracture was approximately 5.6%/y at 5 years and 4.7%/y at 10 years after implantation. The annual risk of protrusion was 1.5%. A total of 40 spontaneous injuries were reported to a worldwide registry (WWR) that included data from 34 672 patients (34 892 Accufix leads), including pericardial tamponade (n=19), pericardial effusion (n=5), atrial perforation (n=3), J retention wire embolization (n=4), and death (n=6). The risk of injury was 0.02%/y (95% CI, 0.0025 to 0. 072) in the MCS and 0.048%/y (95% CI, 0.035 to 0.067) in the WWR. A total of 5299 leads (13%) have been extracted worldwide. After recall in the WWR, fatal extraction complications occurred in 0.4% of intravascular procedures (16 of 4023), with life-threatening complications in 0.5% (n=21). Extraction complications increased with implant duration, female sex, and J retention wire protrusion. CONCLUSIONS Accufix pacing leads pose a low, ongoing risk of injury. Extraction is associated with substantially higher risks, and a conservative management approach is indicated for most patients.
منابع مشابه
Transvenous extraction of a 26-year-old Accufix atrial lead using TightRail rotating dilator sheath.
With the increasing number of implanted pacemakers and implantable cardioverter defibrillators, removal is required more frequently. Presently described is the transvenous extraction of a 26-year-old Accufix atrial lead using a mechanical dilator sheath. A 50-year-old male patient was admitted to the clinic with a pacemaker pocket infection. The atrial lead was an Accufix Bipolar J-Atrial activ...
متن کاملComparison of mid-term clinical experience with steroid-eluting active and passive fixation ventricular electrodes in children.
Although active fixation ventricular leads seem to have advantages over passive fixation leads, this study compares the follow-up results of active and passive fixation leads in children. We evaluated the implantation and follow-up data of 41 children with active (Accufix II DEC, group 1) (n = 20) or passive (Membrane E, group 2) (n = 21) fixation, steroid-eluting ventricular leads. All but one...
متن کاملPercutaneous extraction of a 20-year-old Accufix pacemaker lead complicated by intraoperative protrusion of its J retention wire
Case report A 75-year-old man presented with recurrent persistent Staphylococcus bacteremia from spinal osteomyelitis. He had a past medical history significant for complete heart block for which a pacemaker with an Accufix ventricular lead (model number 330-258) was placed in 1990 and then an Accufix right atrial (RA) lead (model unavailable) in 1995. The bacteremia persisted despite intraveno...
متن کاملHeterotopic Ossification around the Knee after Internal Fixation of a Complex Tibial Plateau Fracture Combined with the Use of Demineralized Bone Matrix (DBM): A Case Report
Demineralized bone matrix has been successfully commercialized as an alternative bone graft material that not only can function as filler but also as an osteoinductive graft. Numerous studies have confirmed its beneficial use in clinical practice. Heterotopic ossification after internal fixation combined with the use of demineralized bone matrix has not been widely reported. In this paper we de...
متن کاملPacemaker lead endocarditis with hiccups (Kalayci)
Background: Lead-related infections that might develop after pacemaker implantation associated with high mortality and morbidity rates are challenging to manage and pose high-cost. Patients with lead-related infections usually present with fever, chills and fatigue and the treatment can be challenging unless the implant system is extracted. Case presentation: A 66-year old male patient who und...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Circulation
دوره 100 23 شماره
صفحات -
تاریخ انتشار 1999