Electronic off-peak status of one head of a dual-head gamma camera resulted in bone scintigraphy artifacts and faulty findings on gated myocardial SPECT.
نویسندگان
چکیده
OBJECTIVE Gamma cameras contain energy discriminators that allow only those photons within a specified energy range to be recorded. A spontaneous shift in peak of 1 head of a dual-head gamma camera may cause artifacts. We present our experience with the incidental occurrence of off-peak status in 1 head of a dual-head gamma camera that resulted in subsequent artifacts and poor-quality images. METHODS Four patients had been scheduled to undergo imaging on a newly installed dual-head gamma camera on the same morning. The first patient underwent (201)Tl-chloride anterior and posterior total-body imaging to check for metastatic thyroid cancer, and the images were of adequate quality for interpretation. The next 2 patients underwent dual-isotope rest (201)Tl-chloride and gated dipyridamole (99m)Tc-tetrofosmin myocardial SPECT. The rest (201)Tl myocardial SPECT images of both patients showed normal perfusion in the left ventricular wall, but the dipyridamole (99m)Tc SPECT images showed virtual absence of perfusion in the apical, anterior, and lateral walls. These findings might suggest myocardial ischemia. In addition, 1 of 2 patients' cardiac gated SPECT findings led to inaccurate left ventricular ejection fractions. Technologists did not become aware of the artifacts until the fourth patient underwent total-body bone scanning, which showed faint activity with loss of contrast in the entire anterior body. One of the camera heads was then found to have off-peak status. Thus, we abandoned use of 1 detector of the dual-head gamma camera and repeated the anterior bone scanning of the 4th patient and the dipyridamole SPECT of 1 of the 2 cardiac patients. RESULTS Gated cardiac SPECT abnormalities resulting from off-peak status were difficult to identify and included abnor-mal left ventricular wall perfusion and an inaccurate left ventricular ejection fraction. It was determined later that the off-peak status was caused by malfunction of a photomultiplier tube. CONCLUSION Degraded planar images resulting from the off-peak status of a camera head are easily identified. The presence of the same pattern of abnormalities in 2 consecutive myocardial SPECT examinations should alert technologists to the possibility that the abnormalities are, in fact, artifacts.
منابع مشابه
Evaluation of myocardial perfusion and function after kidney transplantation by Gated SPECT myocardial perfusion scintigraphy
Introduction: The aim of this study was to evaluate the effect of successful kidney transplantation (KT) on myocardial perfusion and left ventricular function by both qualitative (visual) interpretation and semiquantitative parameters, using myocardial perfusion scintigraphy with gated-single photon emission computed tomography (gated-SPECT) in patients suffering from end-stage renal disease. ...
متن کاملThe role of oral dipyridamole stress gated SPECT in assessing the response of the left ventricle to stress: Re-evaluation of an old method
Introduction: We aimed to assess whether the vasodilator effect of oral dipyridamole on the left ventricular systolic function in patients with suspected CAD is different from that of intravenous (IV) dipyridamole using Tc-99m MIBI myocardial perfusion gated SPECT. Methods: Eighty-nine patients (17 M / 72 W; 61 ± 10 years) were enrolled in this study. The patients underwent a dipyridamole stre...
متن کاملDiastolic left ventricular functional indices: Comparison between gated myocardial perfusion SPECT and echocardiography
Introduction: Many patients have only diastolic dysfunction without any abnormality in systolic function. We compared cardiac diastolic parameters obtained by gated myocardial perfusion SPECT (Gated SPECT) with those evaluated by echocardiography. Methods: Forty-nine patients (aged 37-85 years, 19 males and 30 women) underwent Gated SPECT and echocardiogr...
متن کاملCLINICAL INVESTIGATIONS Poststress Motionlike Artifacts Caused by the Use of a Dual-Head Gamma Camera for 201Tl Myocardial SPECT
When performing 201Tl myocardial SPECT using a dual-head gamma camera on patients after exercise stress, we have observed in some a sudden increase in the counting rate between the 16th and 17th images. This increase provoked motionlike artifacts, which increased the number of false-positive findings. The aim of our study was to determine possible causes for this leap in activity. Methods: We p...
متن کاملBenefits of combined pharmacologic and submaximal exercise stress on Sub-diaphragmatic activity in myocardial perfusion scintigraphy
Introduction: Myocardial perfusion imaging (MPI) is an important imaging modality in managing patients with cardiovascular disease. MPI has a significant role in diagnosis and management of cardiovascular disease; however it is subjected to different artifacts. Combining pharmacologic stress with submaximal exercise reduces side effects, improves image quality, and enhances the...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of nuclear medicine technology
دوره 31 3 شماره
صفحات -
تاریخ انتشار 2003