Intraocular lens power calculation using standard formulas and ray tracing after DMEK in patients with Fuchs endothelial dystrophy

نویسندگان

  • Maged Alnawaiseh
  • Lars Zumhagen
  • André Rosentreter
  • Nicole Eter
چکیده

BACKGROUND The study presented here aims to optimize the accuracy of intraocular lens (IOL) power calculations in patients after DMEK by evaluation of the impact of the altered anterior/posterior corneal curvature relationship. METHODS Scheimpflug-based Oculus Pentacam imaging was performed after DMEK surgery for Fuchs endothelial dystrophy. The IOL power was calculated for all patients by ray tracing, aiming for postoperative emmetropia. We also performed the IOL calculation using four third-generation formulas (SRK-T, Hoffer-Q, Holladay-1 and Haigis). The residual refractions for the individual target IOL were compared and analyzed. RESULTS This retrospective study included 42 eyes of 33 patients (age 68.73 ± 10.11 years) after DMEK surgery. The differences between the expected residual refraction based on ray tracing and that predicted with the third-generation formulas were statistically significant (all formulas p < 0.001). The highest mean difference in the residual refraction between the target IOL measured by ray tracing and that calculated with third-generation formulas was found for the Haigis formula (0.90 ± 0.40 D), and the lowest mean difference for the SRK/T formula (0.73 ± 0.49 D). CONCLUSIONS DMEK surgery induced a relevant change in the anterior to posterior corneal curvature relationship; this needs to be taken into account in the IOL power calculation to avoid hyperopic refractive surprises.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Endothelial keratoplasty for Fuchs dystrophy

We report the case of a 69-year-old female with Fuchs endothelial dystrophy and posterior chamber in-the-bag intraocular lens, whom we treated with DMEK surgical technique. We encountered difficulties both during obtaining the endothelium from the young donor and during the intraocular unrolling and its application on the stroma. We evaluated both preoperative and postoperative the following pa...

متن کامل

Combined Descemet’s membrane endothelial keratoplasty, phacoemulsification and intraocular lens implantation in Fuchs ́dystrophy

METHODS: Prospective, nonrandomized study that included eight eyes of eight patients with concomitant endothelial disease and advanced cataract. In a first group of eight consecutive eyes, combined phacoemulsification, IOL implantation and DMEK were performed. Grafts were inserted with an ICL injector, encapsulated within viscoelastic plugs. Femtosecond laser-assisted descemetorhexis was perfor...

متن کامل

Hydrophilic intraocular lens opacification after posterior lamellar keratoplasty - a material analysis with special reference to optical quality assessment

BACKGROUND Laboratory analysis and optical quality assessment of explanted hydrophilic intraocular lenses (IOLs) with clinically significant opacification after posterior lamellar keratoplasty (DMEK and DSAEK). METHODS Thirteen opacified IOLs after posterior lamellar keratoplasty, 8 after descemet stripping automated endothelial keratoplasty (DSAEK), 3 after descemet membrane endothelial kera...

متن کامل

Intraocular lens calculation for aspheric intraocular lenses.

PURPOSE To evaluate the possible benefits of biometry and ray-tracing intraocular lens (IOL) calculation for aspheric aberration-correcting IOLs. SETTING Private eye clinic in Germany. DESIGN Retrospective consecutive case series. METHODS Eyes with 3 different aberration-correcting IOLs were reviewed. Before surgery, the axial length, corneal thickness, anterior chamber depth, crystalline...

متن کامل

Comparison of OKULIX ray-tracing software with SRK-T and Hoffer-Q formula in intraocular lens power calculation

Purpose To compare the performance of OKULIX ray-tracing software with SRK-T and Hoffer Q formula in intraocular lens (IOL) power calculation in patients presenting with cataract. Methods In this prospective study, 104 eyes of 104 patients with cataract who underwent phacoemulsification and IOL implantation were recruited. Three IOL brands were used and for all eyes, IOL power calculation was...