The purpose of this study was to determine the reproducibility of patient-specific, intensity-modulated radiation therapy (IMRT) quality assurance (QA) results in a clinical setting. Six clinical patient plans were delivered to a variety of devices and analyses, including 1) radiographic film; 2) ion chamber; 3) 2D diode array delivered and analyzed in three different configurations (AP delivery with field-by-field analysis, AP delivery with composite analysis, and planned gantry angle delivery); 4) helical diode array; and 5) in-house-designed multiple ion chamber phantom. The six clinical plans were selected from a range of treatment sites and were of various levels of complexity. Of note, three of the plans had failed at least preliminary evaluation with our in-house IMRT QA; the other three plans had passed QA. These plans were delivered three times sequentially without changing the setup, and then delivered two more times after breaking down and rebuilding the setup between each. This allowed for an investigation of reproducibility (in terms of dose, dose difference or percent of pixels passing gamma) of both the delivery and the physical setup. This study showed that the variability introduced from the setup was generally higher than the variability from redelivering the plan. Radiographic film showed the poorest reproducibility of the dosimeters investigated. In conclusion, the various IMRT QA systems demonstrated varying abilities to reproduce QA results consistently. All dosimetric devices demonstrated a reproducibility (coefficient of variation) of less than 4% in their QA results for all plans, with an average reproducibility of less than 2%. This work provides some quantification for the variability that may be seen for IMRT QA dosimeters.
منابع مشابه
SU-E-T-163: Reproducibility in the Field of Patient-Specific IMRT QA.
PURPOSE To analyze the reproducibility of patient specific IMRT QA results across five different clinical devices and one in-house detector by quantifying variations in the QA results from delivery-to-delivery (without re-setup, "redelivery") and variations when set-up is repeated ("re-setup"). METHODS Six patient plans were selected from among four treatment sites. Six methods of IMRT QA wer...
متن کاملStatistical process control analysis for patient-specific IMRT and VMAT QA
This work applied statistical process control to establish the control limits of the % gamma pass of patient-specific intensity modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) quality assurance (QA), and to evaluate the efficiency of the QA process by using the process capability index (Cpml). A total of 278 IMRT QA plans in nasopharyngeal carcinoma were measured with ...
متن کاملOn the sensitivity of patient‐specific IMRT QA to MLC positioning errors
Accurate multileaf collimator (MLC) leaf positioning plays an essential role in the effective implementation of intensity modulated radiation therapy (IMRT). This work evaluates the sensitivity of current patient-specific IMRT quality assurance (QA) procedures to minor MLC leaf positioning errors. Random errors of up to 2 mm and systematic errors of +/-1 mm and +/-2 mm in MLC leaf positions wer...
متن کاملCharacterization of cylindrical ionization chambers for patient specific IMRT QA
Proven conventional dosimetry techniques do not provide accuracy and precision in the measurement of inverse planned intensity modulated radiation therapy (IMRT) fields. Dynamic and step-and-shoot multileaf collimation (DMLC/SMLC) challenge current ionization chamber dosimetry practices. Ionization chamber performance in these fields is evaluated for three cylindrical chambers of varying volume...
متن کاملDevelopment of a four-axis moving phantom for patient-specific QA of surrogate signal-based tracking IMRT
PURPOSE The purposes of this study were two-fold: first, to develop a four-axis moving phantom for patient-specific quality assurance (QA) in surrogate signal-based dynamic tumor-tracking intensity-modulated radiotherapy (DTT-IMRT), and second, to evaluate the accuracy of the moving phantom and perform patient-specific dosimetric QA of the surrogate signal-based DTT-IMRT. METHODS The four-axi...