Effects of prosthesis surface roughness on the failure process of cemented hip implants after stem-cement debonding.

نویسندگان

  • N Verdonschot
  • E Tanck
  • R Huiskes
چکیده

Retrieval studies suggest that the loosening process of the cemented femoral components of total hip arthroplasties is initiated by failure of the bond between the prosthesis and the cement mantle. Finite element (FE) analyses have demonstrated that stem-cement debonding has stress-producing effects in the cement mantle. High interface friction, which corresponds to a degree of surface roughness, reduces these stresses. In experiments, however, debonded rough stems produced more cement damage than polished ones; in the Swedish Hip Register polished stems were clinically superior with respect to stems with a mat surface finish. The purpose of the present study was to investigate this contradiction. For this purpose, global and local FE models with debonded stem-cement interfaces were used to study the effects of prosthesis surface roughness on the cement stresses on a global scale and microscale, respectively. Similar to earlier numerical studies, the global FE model predicted that an increased surface roughness of the stem reduced the stresses in the cement mantle. The local model provided insight in the load-transfer mechanism on a microscale and could explain the experimental and clinical findings. The local cement peak stresses around the asperities of the surface roughness profile increased with increasing surface roughness and decreased again beyond a particular roughness value. Cement abrasion is caused by localized stresses in combination with micromotion. From this study it can be concluded that to minimize cement abrasion, debonded stems should either have a polished microstructure to minimize the local cement stresses or have a profiled macrostructure to minimize micromotions at the stem-cement interface.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The fixation of the cemented femoral component. Effects of stem stiffness, cement thickness and roughness of the cement-bone surface.

After cemented total hip arthroplasty (THA) there may be failure at either the cement-stem or the cement-bone interface. This results from the occurrence of abnormally high shear and compressive stresses within the cement and excessive relative micromovement. We therefore evaluated micromovement and stress at the cement-bone and cement-stem interfaces for a titanium and a chromium-cobalt stem. ...

متن کامل

Finite element study regarding the influence of the cement mantle thickness on the stress distribution in a cemented hip joint

The disrupted function of the hip joint can be restored by means of total hip arthroplasty (i.e. reconstructive surgery of the joint) THA. This is a widely used medical strategy with high succes rates, world wide [1,2]. Among the failure scenarios of THA, aseptic loosening represents the main reason for revision of cemented hip implants [3]. Studies have shown that the two most frequent ways fo...

متن کامل

Medium-term Outcome of Cemented Titanium Stems

Early failures have seen titanium fall from favour as a material for cemented femoral stems. Between 1989 and 1994, we performed a prospective review of a consecutive series of 122 cemented total hip replacements using the Ultima straight textured titanium stem, and report the fiveto ten-year clinical and radiological outcomes. There were no revisions for loosening of the femoral stem. Revision...

متن کامل

Early aseptic loosening of a mobile-bearing total knee replacement

Background and purpose - Registry-based studies have reported an increased risk of aseptic tibial loosening for the cemented Low Contact Stress (LCS) total knee replacement compared with other cemented designs; however, the reasons for this have not been established. We made a retrieval analysis with the aim of identifying the failure mechanism. Patients and methods - We collected implants, cem...

متن کامل

The effects of cement-stem debonding in THA on the long-term failure probability of cement.

The damage accumulation failure scenario is one of the most prominent ones of cemented THA reconstruction, and involves the accumulation of mechanical damage in materials and interfaces due to repetitive dynamic loading eventually resulting in gross loosening. This study addresses this scenario by combining finite element techniques with the theory of continuum damage mechanics, to analyze the ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of biomedical materials research

دوره 42 4  شماره 

صفحات  -

تاریخ انتشار 1998