Production batch samples of paracetamol tablets and specially prepared out-of-specification batches covering the range 90-110% of the stated amount (500 mg) were analysed by the BP official UV assay and by NIR transmittance spectroscopy. NIR measurements were made on 20 intact tablets from each batch, scanned five times each (10 min measurement time per batch) over the spectral range 6000-11,520 cm-1. An average spectrum was calculated for each batch. Partial least squares (PLS) regression models were set up using a calibration set (20 batches) between the NIR response and the reference tablet paracetamol content (UV). Various pre-treatments of the spectra were examined; the smallest relative standard error of prediction (0.73%) was obtained using the first derivative of the absorbance over the full spectrum. Only two principal components were required for the PLS model to give a good relationship between the spectral information and paracetamol content. Applying this model to the validation set (15 batches) gave a mean bias of -0.08% and a mean accuracy of 0.59% with relative standard deviations of 0.75 and 0.44%, respectively. The proposed method is non-destructive and therefore lends itself to on-line/at-line production control purposes. The method is easy to use and does not require a knowledge of the mass of the tablets.
منابع مشابه
A rapid quantitative assay of intact paracetamol tablets by reflectance near-infrared spectroscopy.
Near-infrared (NIR) reflectance spectroscopy was used to determine rapidly and non-destructively the content of paracetamol in bulk batches of intact Sterwin 500 mg tablets by collecting NIR spectra in the range 1100-2500 nm and using a multiple linear regression calibration method. The developed NIR method gave results comparable to the British Pharmacopoeia 1993 UV assay procedure, the standa...
متن کاملDevelopment of a rapid process monitoring method for dry-coated tableting process by using near-infrared spectroscopy.
A nondestructive transmittance near-infrared (NIR) method for detecting off-centered cores in dry-coated (DC) tablets was developed as a monitoring system in the DC tableting process. Caffeine anhydrate was used as a core active pharmaceutical ingredient (API), and DC tablets were made by the direct compression method. NIR spectra were obtained from these intact DC tablets using the transmittan...
متن کاملDetermining Blood Glucose Concentration using Near Infrared Spectroscopy: Early Findings
Introduction: Diabetes mellitus is one of the diseases that have grown dramatically in today's societies. People with diabetes should continuously measure their blood glucose level. Continuous blood glucose measurement by commonly used methods is painful and difficult. On the other hand, mobile phone can be a useful tool for accessing physicians and telemedicine services more easily. The aim of...
متن کاملDetermining Blood Glucose Concentration using Near Infrared Spectroscopy: Early Findings
Introduction: Diabetes mellitus is one of the diseases that have grown dramatically in today's societies. People with diabetes should continuously measure their blood glucose level. Continuous blood glucose measurement by commonly used methods is painful and difficult. On the other hand, mobile phone can be a useful tool for accessing physicians and telemedicine services more easily. The aim of...
متن کاملNondestructive Firmness Estimation of Tomato Fruit Using Near-Infrared Spectroscopy
Today, nondestructive methods are widely used to determine the quality of agricultural products. Meanwhile, visible and near-infrared (Vis/NIR) spectroscopy is regarded as one of the most widely used methods in the field of quality assessment of agricultural products. In this study, a system was developed to measure the Vis/NIR spectra of tomato fruit samples in the half-transmittance mode of m...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Analyst
دوره 123 11 شماره
صفحات -
تاریخ انتشار 1998