QTc Heterogeneity in Rest Magnetocardiography is Sensitive to Detect Coronary Artery Disease: In Comparison with Stress Myocardial Perfusion Imaging.
نویسندگان
چکیده
BACKGROUND Stress nuclear myocardial perfusion imaging (MPI) is an established method for diagnosis and prognosis of coronary artery disease (CAD). However, radiation exposure limits its clinical application. Magnetocardiography (MCG) has been proposed as a non-contact, rapid and non-radiation technique with high reproducibility. The aim of the study was to evaluate the diagnostic efficacy of rest MCG in CAD comparing to stress MPI. METHODS We prospectively enrolled 55 patients with suspected CAD (64 ± 10 years) who were scheduled for coronary angiography (CA). MCG, stress (201)Tl MPI and CA were performed within 3 months. The spatial distribution maps of QTc interval (21 × 21 in resolution) were derived from a 64-channel MCG system (KRISS, Korea). T-wave propagation mapping, repolarization heterogeneity index with QTc dispersion and smoothness index of QTc (SI-QTc) were analyzed, and the diagnostic criteria for CAD were developed based on the receiver operating characteristic (ROC) curve analysis. RESULTS Patients with significant CAD (≥ 70% luminal stenosis, n = 36) had higher QTc dispersion and SI-QTc than controls (both p < 0.05). The diagnostic sensitivity and specificity were 0.8330, 0.6842 for QTc dispersion ≥ 79 ms; 0.7778, 0.6842 for SI-QTc ≥ 9.1 ms; and 0.8611, 0.6842 for combination. There was no difference of area under ROC curve by using criteria of QTc dispersion ≥ 79 ms, SI-QTc ≥ 9.1 ms or combination (0.7588, 0.7310, 0.7727, p = NS), and non-inferior to stress MPI (p = NS). CONCLUSIONS The QTc heterogeneity parameters of rest MCG yield a good sensitivity and acceptable specificity for detection of CAD, and may provide an alternative to stress MPI without stress and radiation. KEY WORDS Coronary artery disease (CAD); Magnetocardiography (MCG); Myocardial perfusion imaging (MPI); Repolarization.
منابع مشابه
Myocardial perfusion scan accuracy in detection of coronary artery disease - Comparison with exercise stress test [Persian]
Introduction: In patients with coronary artery disease (CAD) noninvasive evaluation for detection of ischemia is important to avoid invasive interventions like angiography. Exercise stress test is conventionally the first study used in evaluation of CAD. Considering the noninvasive nature of the myocardial perfusion scan, we decided to compare its accuracy with stress test. Methods: Patie...
متن کاملComparison of dipyridamole – induced electrocardiographic changes and myocardial perfusion imaging in diagnosis of coronary artery disease [Persian]
Introduction: This study evaluated the diagnostic value of the dipyridamole induced electrocardiuogram (ECG) cahnges, comparing with myocardial perfusion imaging. Materials and Methods: 222 patinets were studied with dipyridamole infusion (based on ECG criteria) as well as Dipyridamole-myocardial perfusion imaging (MPI). Results: Abnormal dipyridamole test and abnormal MPI were noted in 1...
متن کاملGated 99mTc-MIBI myocardial perfusion SPECT in patients with right bundle branch block but without evidence of coronary artery disease
Introduction: We aimed to investigate the effect of right bundle branch block (RBBB) on perfusion and functional parameters in dipyridamole stress/rest Tc99m-MIBI gated myocardial perfusion SPECT (GSPECT) which may be helpful in interpretation of myocardial perfusion imaging. Methods: We studied 73 patients with low pre-test likelihood of coronary artery disease in two groups: 38 patients wit...
متن کاملTwo-day stress-rest lower limbs perfusion scan in patients referred for myocardial perfusion imaging
Introduction: Peripheral Vascular Disease (PVD) is a major cause of morbidity and is associated with Coronary Artery Disease (CAD). We aimed to perform Lower Limb Perfusion Scan (LLPS) in patients referred for Myocardial Perfusion Imaging (MPI) and estimate prevalence of PVD in subgroups with normal and abnormal MPI results. We also compared quantitative indices of LLPS in pati...
متن کاملEarly Myocardial Repolarization Heterogeneity Is Detected by Magnetocardiography in Diabetic Patients with Cardiovascular Risk Factors
Multi-channel magnetocardiography (MCG) is a sensitive technique to map spatial ventricular repolarization with high resolution and reproducibility. Spatial ventricular repolarization heterogeneity measured by MCG has been shown to accurately detect and localize myocardial ischemia. Here, we explored whether these measurements correlated with cardiovascular risk factors in patients with type 2 ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Acta Cardiologica Sinica
دوره 30 5 شماره
صفحات -
تاریخ انتشار 2014