Wear of a composite ceramic head caused by liner fracture.

نویسندگان

  • Michael M Morlock
  • Florian Witt
  • Nick Bishop
  • Rainer Behn
  • Paolo Dalla Pria
  • Rob Barrow
  • Ian Dymond
چکیده

Third-generation composite ceramics (eg, Delta; DePuy Orthopaedics, Warsaw, Indiana; Ceramtec, Plochingen, Germany) have greatly improved material characteristics compared with second-generation products. This case report presents a patient after total hip arthroplasty with a fractured ceramic liner and a heavily worn ceramic head (both third-generation ceramics) retrieved 9 months after surgery. The patient showed no symptoms in the involved hip but presented to the hospital because of other symptoms. The failure was caused by a tilted liner that was overlooked after surgery and fractured consecutively. Rim chipping and splitting were the 2 fracture modes observed for the liner. The head did not fracture completely because of its high strength but became roughened by the ceramic fragments, causing major wear of the metal back of the cup. The phase transformation of the zirconium grains from tetragonal to monoclinic in the aluminum oxide matrix was shown by radiographic diffraction analysis in the heavily worn areas of the head. This transformation increases the fracture strength of the head. Metal debris caused by a roughened ceramic head without fracture is an unreported phenomenon for third-generation ceramic bearings in hip arthroplasty. This case shows that proper impaction of the ceramic liner into the metal shell to prevent later tiling during reduction is as important as correct component positioning. If a tilted ceramic liner is observed postoperatively, the position must be corrected immediately to prevent the major consequences observed in this patient.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Repeated Ceramic Head Fracture after Ceramic-on-ceramic Total Hip Arthroplasty

Early generation ceramic articulations were problematic in terms of ceramic implant fracture and wear. Although the mechanical properties of ceramic material have improved during the last three decades, ceramic component fracture remains an issue. Many authors have reported fractures of the ceramic head, and a few have described ceramic acetabular insert fracture. However, these complications a...

متن کامل

Ceramic Head Fracture in Ceramic-on-Polyethylene Total Hip Arthroplasty

Revision rates of total hip arthroplasty have decreased after introducing total hip arthroplasty (THA) using ceramic component, since ceramic components could reduce components wear and osteolysis. The fracture of a ceramic component is a rare but potentially serious event. Thus, ceramic on polyethylene articulation is gradually spotlighted to reduce ceramic component fracture. There are a few ...

متن کامل

Adverse local tissue reaction after 2 revision hip replacements for ceramic liner fracture

INTRODUCTION In younger patients, ceramic-on-ceramic (CoC) bearing surfaces are usually recommended for total hip replacement (THR) because of their low wear rate and longer expected functional life. Although technical advancements have reduced the risk of ceramic bearings fracture, this complication remains a major concern. CASE DESCRIPTION We present the case of a 56-year-old patient underg...

متن کامل

Fracture of Ceramic Liner and Head in a Total Hip Arthroplasty with a Sandwich Type Cup

Due to its advantages, ceramic-on-ceramic bearings have been widely used in young patients for almost 30 years. Long-term survivorship, low wear, and low biological reactivity to particles are some of its characteristics. Even though this material has had a lot of improvements, the risk of fracture is one of the concerns. There have been reports of fracture of ceramic in the acetabular liner an...

متن کامل

Severe metallosis leading to femoral head perforation.

This article describes a case of severe metallosis in a 67-year-old woman who initially underwent primary total hip arthroplasty with a ceramic-on-ceramic articular bearing. This was subsequently revised to a metal-on-polyethylene articulation due to ceramic liner fracture. She presented with severe hip pain and a pelvic mass. Infective workup was negative. Perforation of the cobalt-chrome femo...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Orthopedics

دوره 37 7  شماره 

صفحات  -

تاریخ انتشار 2014