An audit of rejected repeated x-ray films as a quality assurance element in a radiology department.

نویسندگان

  • K C Eze
  • N Omodia
  • B Okegbunam
  • T Adewonyi
  • C C Nzotta
چکیده

OBJECTIVES To find out the causes, number, percentage and sizes of rejected radiographic films with a view of adopting measures that will reduce the rate and number of rejected films. SETTING Radiology Department of a University Teaching Hospital. MATERIALS AND METHODS Over a two-year period (1st April 2002 to 31st March 2004), the total number of x-ray films utilized for radiographic examinations, rejected films and sizes of rejected films were collected retrospectively from the medical record of radiology department. All the rejected films were viewed by a radiologist and three radiographers for the causes of the rejects which was arrived at by consensus. The data was analysed. RESULT A total of 15,095 films were used in the study period and 1,338 films (8.86%) were rejected or wasted. The rate of rejected films varied from 7.69% to 13.82% with average of 8.86%. The greatest cause of film rejects was radiographers' faults 547 (40.88%), followed by equipments faults 255 (19.06%), and patients' faults 250 (18.90%). The highest reject rate (13.82%) was for films used for examination of the spine (15 x 30) cm size. This is followed by 9.92% for skull (18 x 24) cm films and 8.83% for small sized films (24 x 30) cm used for paediatric patients. Of a total of 1,338 rejected films, 1276 (95.37%) additional exposure were done to obtain the basic desired diagnostic information involving 1151 patients; 885 (76.89%) of these patients needed at least one additional hospital visit to take the repeat exposure. CONCLUSION Rejected films are not billable; patients receive additional radiation and may even come to hospital in another day for the repeat. Radiographer's work is increased as well as that of the support staff. The waiting room may be congested and waiting time increased. The cost of processing chemical and films are increased, thus if work is quantified in monetary terms, the cost of repeats is high. Rejected-repeated film analysis is cheap, simple, practicable, easy to interpret and an effective indictor of quality assurance of radiology departments.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Analysis and Economic Implication of X-Ray Film Reject in Diagnostic Radiology Department of Jimma University Specialized Hospital, Southwest Ethiopia

Background Patients usually undergo repeated X-ray examinations after their initial X-ray radiographs are rejected due to poor image quality. This subjects the patients to excess radiation exposure and extra cost.It is therefore investigating the magnitude and causes of reject is mandatory. This study aimed to assess the reject rate of X-ray films and its economic implication in order to obtain...

متن کامل

New Method of Quality Control Test for Light and Radiation Field Coincidence in Medical Linear Accelerators

Introduction: The evaluation of X-ray and light field coincidence in linear accelerators as a quality control test is often performed subjectively, involving the manual marking of films and their visual inspection following the irradiation. Therefore, the present study aimed to develop an objective method for the performance of this test leading to the increased levels...

متن کامل

Film Reject Analysis and Radiation Doses Received by Patients in Selected Hospitals in Southwestern Nigeria

Introduction: A reject rate is the percentage of diagnostic images repeated due to errors during radiological examinations. The present study aimed to evaluate the patient radiation doses and analyze the film reject rate as part of quality assurance program in three diagnostic centers in Nigeria. Materials and Methods: This study was conducted in three hospitals, namely Federal Medical Center ...

متن کامل

Evaluation of the Quality Control Program for Diagnostic Radiography and Fluoroscopy Devices in Syria during 2005-2013

Introduction: Extensive use of diagnostic radiology is the largest contributor to total population radiation doses. Thus, appropriate equipment and safe practice are necessary for good-quality images with optimal doses. This study aimed to perform quality control (QC) audit for radiography and fluoroscopy devices owned by private sector in Syria (2005-2013) to verify compliance of performance o...

متن کامل

New electrical device for field coincidence quality control test in linear accelerators

Testing of the coincidence of a linear accelerators X-ray field and the light field as a quality control test is often done in a subjective method, involving the manual marking of pieces of film and the visual inspection of the film after irradiation. The purpose of this study was to develop an objective method for performing this test, while also increasing the accuracy, precision and the spee...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Nigerian journal of clinical practice

دوره 11 4  شماره 

صفحات  -

تاریخ انتشار 2008