PVC Membrane Sensors for Potentiometric Determination of Acebutolol
نویسندگان
چکیده
The construction and general performance characteristics of two novelpotentiometric membrane sensors responsive to the acebutolol are described. Thesensors are based on the use of ion-association complexes of acebutolol (AC) withtetraphenylborate(TPB) (I) and phosphomolybdate(PM) (II) as exchange sites in a PVCmatrix. The sensors show a fast, stable and near- Nernstian for the mono charge cationof AC over the concentration range 1×10-3 - ~10-6 M at 25 °C over the pH range 2.0 -6.0 with cationic slope of 51.5 ± 0.5 and 53.0 ± 0.5 per concentration decade for AC-Iand AC-II sensors respectively. The lower detection limit is 6×10-6 M and 4×0-6 M withthe response time 20-30 s in the same order of both sensors. Selectivity coefficients ofAC related to a number of interfering cation and some organic compounds wereinvestigated. There are negligible interferences are caused by most of the investigatedspecies. The direct determination of 3 - 370 μg/ml of AC shows an average recovery of 99.4 and 99.5% and a mean relative standard deviation of 1 . 5 % at 100.0 μg/ml forsensor I and II respectively. The results obtained by determination of AC in tablets usingthe proposed sensors which comparable favorably with those obtained by the Britishpharmacopoeia method. In the present investigation the electrodes have been utilized asend point indicator for some precipitation titration reactions.
منابع مشابه
Green Potentiometric Method for Determination of Triprolidine Hydrochloride and Pseudoephedrine Hydrochloride in their Different Pharmaceutical Matrices using Liquid and Solid Contact Gold Electrodes
Factors affecting the performance of polyvinyl chloride based ion selective electrode such as using different cationic exchangers and ionophores as well as sensor fabrication procedures were recently the scope of study by several researchers. This work presents an environment friendly electro-analytical method for the determination of Triprolidine HCl (TRI) and Pseudoephedrine HCl (PSE). A comp...
متن کاملApplication of a dioxime-PVC electrode to potentiometric studies of Cr(III) ion
A PVC membrane Cr(III) ion selective electrode has been constructed using 2H-1,4-benzothioazine-2,3(4H)dione dioxime (BTD) as membrane carrier. The influence of membrane composition on the electroderesponse was studied. The electrode exhibits a Nernstian response over a Cr(III) concentration range of 1.0 x10 6 to 1.0>< 10-1 M (r = 0.99) with a slope of 19.5 ± 0.5 mV per decade of concentration,...
متن کاملPerformance Evaluation of a Novel Potentiometric Membrane Sensor for Determination of Atorvastatin in Pharmaceutical Preparations
A novel potentiometric ion-selective PVC membrane sensor for analysis of atorvastatin (AT) in pharmaceutical preparations based on atorvastatin-(tetraphenyl borate), (AT-(TPB)2) as sensing element, tetraphenyl borate as additive and tris-2-ethyl-hexyl phosphate (TOP) as plasticizer solvent was prepared. The electrode shows a good Nernestian response over the concentration range of 0.09–5586 µ...
متن کاملPerformance Evaluation of a Novel Potentiometric Membrane Sensor for Determination of Atorvastatin in Pharmaceutical Preparations
A novel potentiometric ion-selective PVC membrane sensor for analysis of atorvastatin (AT) in pharmaceutical preparations based on atorvastatin-(tetraphenyl borate), (AT-(TPB)2) as sensing element, tetraphenyl borate as additive and tris-2-ethyl-hexyl phosphate (TOP) as plasticizer solvent was prepared. The electrode shows a good Nernestian response over the concentration range of 0.09–5586 µ...
متن کاملPotentiometric Determination of Ketotifen Fumarate in Pharmaceutical Preparations and Urine Using Carbon Paste and PVC Membrane Selective Electrodes
This study compares between unmodified carbon paste (CPE; the paste has no ion pair) and polyvinyl chloride (PVC) membrane selective electrodes that were used in potentiometric determination of ketotifen fumarate (KTF), where sodium tetraphenylborate (NaTPB) was used as titrant. The performance characteristics of these sensors were evaluated according to IUPAC recommendations which reveal a fas...
متن کامل