Epidural catheter penetration of human dural tissue: in vitro investigation.

نویسندگان

  • Pamela J Angle
  • Jean E Kronberg
  • Dorothy E Thompson
  • James Duffin
  • Peter Faure
  • Shan Balasubramaniam
  • John Paul Szalai
  • Susan Cromwell
چکیده

BACKGROUND Factors contributing to subarachnoid catheter passage after epidural placement are not well understood. This study explored mechanisms that might explain its occurrence. METHODS Human cadaveric dura was mounted on a model and pressurized to physiologic levels. In a standardized fashion, a 20-gauge Portex three-port, closed end (nonflexible) tip catheter was passed through an epidural needle mounted on a micromanipulator at a 90 degree angle, attempting to penetrate dura with the catheter. Attempts then followed with a 19-gauge Arrow Flex Tip Plus single-port catheter. Subarachnoid catheter passage was compared in (1) intact dura, (2) clinically occult versus obvious epidural needle punctures, and (3) single 25-gauge Whitacre spinal needle punctures after combined spinal-epidural placement. RESULTS Neither catheter penetrated intact dura: Portex, 0 of 300 attempts (0.0000; 95% confidence interval [CI]: 0.0000, 0.0158); Arrow, 0 of 300 attempts (0.0000; 95% CI: 0.0000, 0.0158). In clinically occult epidural needle punctures, the 20-gauge Portex catheter penetrated 1 of 3 specimens in 1 of 14 attempts (0.0714; 95% CI: 0.0021, 0.3583). The 19-gauge Arrow did not pass (0 of 15 attempts, 0.0000; 95% CI: 0.0000, 0.2535). In clinically obvious epidural needle punctures, the Portex passed in 6 of 33 attempts (0.1818; 95% CI: 0.0760, 0.3608) and the Arrow passed in 1 of 35 attempts (0.0286; 95% CI: 0.0012, 0.1662). Neither catheter passed through a single 25-gauge spinal needle puncture after an uncomplicated combined spinal-epidural: Portex, 0 of 90 attempts (0.0000; 95% CI: 0.0000, 0.0510); Arrow, 0 of 90 attempts (0.0000; 95% CI: 0.0000, 0.0510). CONCLUSIONS Catheter passage is unlikely in the presence of intact dura or after an uncomplicated combined spinal-epidural. Unintentional subarachnoid passage suggests dural damage with the epidural needle.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

بی‌حسی کامل نخاعی در بیمار مبتلا به آکندروپلازی: گزارش موردی

Background: Total spinal anesthesia is a complication of lumbar epidural anesthesia following undiagnosed subarachnoid or subdural injection of local anesthetic. Although many achondroplastic dwarfs have a normal spine, catheter insertion may be more problematic with a narrow epidural space making a subarachnoid tap more probable.  Other malformations associated with achondroplasia, such as pro...

متن کامل

Decrease in the incidence of post-dural puncture headache: maintaining CSF volume.

The incidence of epidural needle-induced post-dural puncture headache (PDPH) in parturients following dural puncture with a large bore (18-gauge) needle has been reported to range 76-85%. We describe seven cases in which the performance of epidural anesthesia in parturients was complicated by an unintentional dural puncture with an 18-gauge epidural needle. In all seven cases, the unintentional...

متن کامل

An alternative management procedure after inadvertent dural puncture

BACKGROUND Accidental dural puncture during epidural blockade is a rare complication. Since postdural puncture headache (PDPH) is the most frequent complication, a wisely management method will also reduce the incidence of PDPH. METHODS Five patients who had inadvertent dural puncture during the epidural procedure before hip or knee arthroplasty were included in this study. After dural punctu...

متن کامل

Insertion of an intrathecal catheter in parturients reduces the risk of post-dural puncture headache: A retrospective study and meta-analysis

This study aimed to determine whether insertion of an intrathecal catheter following accidental dural puncture (ADP) in obstetric patients can reduce the incidence of post-dural puncture headache (PDPH) and the requirement of a therapeutic epidural blood patch (TEBP). This was also compared with relocating the epidural catheter at a different vertebral interspace. A retrospective study was perf...

متن کامل

Nonpharmacologic Neuraxial Interventions for Prophylaxis of Postdural Puncture Headache in the Obstetric Patient.

Postdural puncture headache due to accidental dural puncture during epidural catheter placement is a source of morbidity for new mothers. It can interfere with maternal-newborn bonding and increase the length of hospitalization. This evidence-based article examined the question: For obstetric patients experiencing an accidental dural puncture during epidural placement, which nonpharmacologic pr...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Anesthesiology

دوره 100 6  شماره 

صفحات  -

تاریخ انتشار 2004