Comparison of serious inhaler technique errors made by device-naïve patients using three different dry powder inhalers: a randomised, crossover, open-label study
نویسندگان
چکیده
BACKGROUND Serious inhaler technique errors can impair drug delivery to the lungs. This randomised, crossover, open-label study evaluated the proportion of patients making predefined serious errors with Pulmojet compared with Diskus and Turbohaler dry powder inhalers. METHODS Patients ≥18 years old with asthma and/or COPD who were current users of an inhaler but naïve to the study devices were assigned to inhaler technique assessment on Pulmojet and either Diskus or Turbohaler in a randomised order. Patients inhaled through empty devices after reading the patient information leaflet. If serious errors potentially affecting dose delivery were recorded, they repeated the inhalations after watching a training video. Inhaler technique was assessed by a trained nurse observer and an electronic inhalation profile recorder. RESULTS Baseline patient characteristics were similar between randomisation arms for the Pulmojet-Diskus (n = 277) and Pulmojet-Turbohaler (n = 144) comparisons. Non-inferiority in the proportions of patients recording no nurse-observed serious errors was demonstrated for both Pulmojet versus Diskus, and Pulmojet versus Turbohaler; therefore, superiority was tested. Patients were significantly less likely to make ≥1 nurse-observed serious errors using Pulmojet compared with Diskus (odds ratio, 0.31; 95 % CI, 0.19-0.51) or Pulmojet compared with Turbohaler (0.23; 0.12-0.44) after reading the patient information leaflet with additional video instruction, if required. CONCLUSIONS These results suggest Pulmojet is easier to learn to use correctly than the Turbohaler or Diskus for current inhaler users switching to a new dry powder inhaler. TRIAL REGISTRATION ClinicalTrials.gov Identifier: NCT01794390 (February 14, 2013).
منابع مشابه
Predictors of incorrect inhalation technique in patients with asthma or COPD: a study using a validated videotaped scoring method.
BACKGROUND Inadequate technique reduces the effects of inhalation medication. Errors in inhalation technique have been reported to range up to 85%. Not only various patients' characteristics but also the device has an effect on correct inhalation technique. The purpose of this study was to determine the effect of patients' characteristics and type of inhaler device on inhalation technique in pa...
متن کاملEvaluation of inhaler handling-errors, inhaler perception and preference with Spiromax, Easyhaler and Turbuhaler devices among healthy Finnish volunteers: a single site, single visit crossover study (Finhaler).
INTRODUCTION Correct inhaler technique and device preference are positively correlated with improved adherence and clinical outcomes. This study was designed to investigate inhaler technique mastery and device preference for three different dry powder inhalers, Spiromax, Easyhaler and Turbuhaler. METHODS This was a single site, single visit, crossover study assessing device mastery, handling ...
متن کاملMultiple inhalers confuse asthma patients.
This study investigated the influence of the use of different types of inhalers on the adequacy of inhalation technique among adult asthmatics. Three hypotheses were tested: first, patients using only one type of inhaler will demonstrate adequate inhalation technique more often than those with two or more types. Secondly, patients using a combination of dry powder inhalers (DPIs) will demonstra...
متن کاملEvaluation of the effectiveness of four different inhalers in patients with chronic obstructive pulmonary disease.
BACKGROUND The percentage of patients inhaling their medication effectively varies widely, according to methods of assessment and inhalers used. This study was carried out to assess differences among four types of inhalers using inhaler-specific checklists. METHODS Inhalation technique was evaluated in adult patients with chronic obstructive pulmonary disease (COPD). Inhalers investigated wer...
متن کاملComparison of a new multidose powder inhaler (Diskus/Accuhaler) and the Turbuhaler regarding preference and ease of use.
Many patients with asthma or chronic obstructive pulmonary disease (COPD) use their inhaler ineffectively. Several studies revealed that inhaler-specific design features contributed significantly to the failure rate, which clearly demonstrates the need for developing more "foolproof" inhalers. This study compared ease of use and patient preference of the Diskus/Accuhaler (DA) with the Turbuhale...