In vivo subjective and objective longitudinal chromatic aberration after bilateral implantation of the same design of hydrophobic and hydrophilic intraocular lenses.

نویسندگان

  • Maria Vinas
  • Carlos Dorronsoro
  • Nuria Garzón
  • Francisco Poyales
  • Susana Marcos
چکیده

PURPOSE To measure the longitudinal chromatic aberration in vivo using psychophysical and wavefront-sensing methods in patients with bilateral implantation of monofocal intraocular lenses (IOLs) of similar aspheric design but different materials (hydrophobic Podeye and hydrophilic Poday). SETTING Instituto de Optica, Consejo Superior de Investigaciones Cientificas, Madrid, Spain. DESIGN Prospective observational study. METHODS Measurements were performed with the use of psychophysical (480 to 700 nm) and wavefront-sensing (480 to 950 nm) methods using a custom-developed adaptive optics system. Chromatic difference-of-focus curves were obtained from best-focus data at each wavelength, and the longitudinal chromatic aberration was obtained from the slope of linear regressions to those curves. RESULTS The longitudinal chromatic aberration from psychophysical measurements was 1.37 diopters (D) ± 0.08 (SD) (hydrophobic) and 1.21 ± 0.08 D (hydrophilic). From wavefront-sensing, the longitudinal chromatic aberration was 0.88 ± 0.07 D and 0.73 ± 0.09 D, respectively. At 480 to 950 nm, the longitudinal chromatic aberration was 1.27 ± 0.09 D (hydrophobic) and 1.02 ± 0.13 D (hydrophilic). The longitudinal chromatic aberration was consistently higher in eyes with the hydrophobic IOL than in eyes with the hydrophilic IOL (a difference of 0.16 D and 0.15 D, respectively). Similar to findings in young phakic eyes, the longitudinal chromatic aberration from the psychophysical method was consistently higher than from wavefront-sensing, by 0.48 D (35.41%) for the hydrophobic IOL and 0.48 D (39.43%) for the hydrophilic IOL. CONCLUSION Longitudinal chromatic aberrations were smaller with hydrophilic IOLs than with hydrophobic IOLs of the same design. FINANCIAL DISCLOSURE No author has a financial or proprietary interest in any material or method mentioned.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparing the effect of two different intraocular lenses on optical aberrations in bilaterally operated eyes for cataract

OBJECTIVE To assess high order and spherical aberrations results of hydrophobic acrylic AMO Sensar AR40E and hydrophobic acrylic Alcon AcrySof SA60AT intraocular lenses after implantation in cases with bilateral cataract. METHODS Cases diagnosed as bilateral cataract were included in the study and preoperative aberration measurements were recorded by using Nidek OPD SCAN-ARK 1000. Groups were...

متن کامل

Comparison of hydrophobic and hydrophilic intraocular lens in preventing posterior capsule opacification after cataract surgery

BACKGROUND Posterior capsular opacification (PCO) is a common long-term complication of cataract surgery. Intraocular lens design and material have been implicated in influencing the development of PCO. This study evaluated the association of hydrophobic and hydrophilic intraocular lenses on preventing PCO. METHODS Medline, Cochrane, EMBASE, and Google Scholar databases were searched until Au...

متن کامل

Neodymium: YAG capsulotomy rates after phacoemulsification with hydrophobic and hydrophilic acrylic intraocular lenses.

PURPOSE To retrospectively evaluate the incidence of neodymium: YAG (ND:YAG) laser capsulotomies for posterior capsule opacification (PCO) seen with hydrophobic and hydrophilic acrylic intraocular lens implants 18 months after surgery. METHODS One hundred and five eyes that had phacoemulsification with a continuous curvilinear capsulorhexis and implantation of either a hydrophobic or a hydrop...

متن کامل

Findings of an experimental study in a rabbit model on posterior capsule opacification after implantation of hydrophobic acrylic and hydrophilic acrylic intraocular lenses

PURPOSE Study on cell growth on the posterior capsule after implantation of hydrophobic acrylic (Acrysof SA 60 AT) and hydrophilic acrylic (Akreos Disc) intraocular lenses (IOL) in a rabbit model and comparison of posterior capsule opacification (PCO). METHODS Phacoemulsification was performed in 22 rabbit eyes, and two different IOL types (Acrysof SA60 AT and Akreos Disc) were implanted. The...

متن کامل

Achromatic doublet intraocular lens for full aberration correction.

A doublet intraocular lens optimized for both chromatic and monochromatic aberration correction in pseudophakic eyes is presented. Ray-tracing techniques were applied to design the lens in white light within a chromatic eye model. Combinations of two materials, already commonly used in intraocular lenses, as acrylic and silicone, were used. Iterative optimization algorithms were employed to cor...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of cataract and refractive surgery

دوره 41 10  شماره 

صفحات  -

تاریخ انتشار 2015