The clinical impact of the couch top and rails on IMRT and arc therapy.
نویسندگان
چکیده
The clinical impact of the Varian Exact Couch on dose, volume coverage to targets and critical structures, and tumor control probability (TCP) has not been described. Thus, we examined their effects on IMRT and arc therapy. Five clinical prostate patients were planned with both 6 MV eight-field IMRT and 6 MV two-arc RapidArc techniques using the Eclipse treatment planning system. These plans neglected treatment couch attenuation, as is a common clinical practice. Dose distributions were then recalculated in Eclipse with the inclusion of the Varian Exact Couch (imaging couch top) and the rails in varying configurations. The changes in dose and coverage were evaluated using the dose-volume histograms from each plan iteration. We used a TCP model to calculate losses in tumor control resulting from not accounting for the couch top and rails. We also verified dose measurements in a phantom. Failure to account for the treatment couch and rails resulted in clinically unacceptable dose and volume coverage losses to the targets for both IMRT and RapidArc. The couch caused average prescription dose losses (relative to plans that ignored the couch) to the prostate of 4.2% and 2.0% for IMRT with the rails out and in, respectively, and 3.2% and 2.9% for RapidArc with the rails out and in, respectively. On average, the percentage of the target covered by the prescribed dose dropped to 35% and 84% for IMRT (rails out and in, respectively) and to 18% and 17% for RapidArc (rails out and in, respectively). The TCP was also reduced by as much as 10.5% (6.3% on average). Dose and volume coverage losses for IMRT plans were primarily due to the rails, while the imaging couch top contributed most to losses for RapidArc. Both the couch top and rails contribute to dose and coverage losses that can render plans clinically unacceptable. A follow-up study we performed found that the less attenuating unipanel mesh couch top available with the Varian Exact couch does not cause a clinically impactful loss of dose or coverage for IMRT but still causes an unacceptable loss for RapidArc. Therefore, both the imaging or mesh couch top and the rails should be accounted for in arc therapy. The imaging couch top should be accounted for in IMRT treatment planning or the mesh top can be used, which would not need to be accounted for, and the rails should be moved to avoid the beams during treatment.
منابع مشابه
Characterization of dose impact on IMRT and VMAT from couch attenuation for two Varian couches
In intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT), the use of posterior oblique beams has become common. Beam attenuation by the treatment couch is not negligible when the couch is in the beam portal. In this study, we established the relationship of relative dose vs. beam angle for two Varian 21EX linacs, one equipped with the Exact couch (standard cou...
متن کاملQuantification and modelling of the dosimetric impact of the treatment couch in volumetric modulated arc therapy (VMAT)
Background: As the volumetric modulated arc therapy (VMAT) becoming a main role of treatment ways, the effect of couch top becomes more significant. It is imperative to re-evaluate the couches that previously may have been considered of no importance during early treatment techniques. The impact of couch top on radiation delivery was explored and the couch model was tested with the aim of reduc...
متن کاملSU-E-T-549: On Importance of Accurate Treatment Couch Modeling in SmartArc VMAT Plans.
PURPOSE We investigate the impact of a treatment couch and its modeling in a treatment planning system on accurate dose delivery in Volumetric Modulated Arc Therapy (VMAT). METHODS A cone-beam CT of a Varian couch on a Varian Clinac 21EX unit, consisting of a carbon-fiber imaging couch top and supporting rails, was acquired, and a couch model was created in Philips Pinnacle 9.0 TPS based on t...
متن کاملDosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...
متن کاملDose Distribution Analysis of Rapid Arc and Intensity Modulated Radiotherapy Plan in Head and Neck Cancer
Introduction: Globally, intensity-modulated radiation therapy (IMRT) is considered as highly precise and accurate method of radiotherapy planning. This technique amplifies spatial dose distribution conformity by modulating the intensity of radiation beams in each sub-volume. Additionally, it can reduces the dose to surrounding critical organs and deliver the planned dose to targets with the nom...