GC/MS Analysis of PCBs in Waste Oil Using the Backflush Capability of the Agilent QuickSwap Accessory Application
نویسنده
چکیده
Polychlorinated biphenyls (PCBs) in waste oil are typically analyzed by GC-ECD or GC/MS after solid phase extraction (SPE) cleanup. However, not all problematic matrix components are completely removed during cleanup and are injected into the analytical system, thereby contaminating the column and the detector. In this application, a practical example of backflushing is presented using the Agilent QuickSwap accessory installed on a 7890 GC/5975 MSD system. Benefits of using QuickSwap instead of the traditional high-temperature bakeout procedure are demonstrated. Column and detector contamination were significantly reduced and sample throughput increased.
منابع مشابه
Automated Clean-up of PCB extracts from Waste Oil using the Agilent 7696A Sample Prep WorkBench
Solid waste, including petroleum based waste oils, transformer oil or mineral oil are screened for the presence of polychlorinated biphenyls (PCBs). The analysis is typically performed by GC-ECD or GC-MS after clean-up using column chromatography, solid phase extraction (SPE), or dispersive solid phase extraction (d-SPE). Clean-up on silica or acidified silica removes polar constituents that ca...
متن کاملتأثیر بافت و رطوبت خاک در استخراج PCBs
PCBs are persistent organic pollutants which, due to high environmental hazards, must be traced, determined, and decomposed to reduce their risks. . To detect this material in the soil, the method of extraction and appropriate measuring conditions should be investigated. Two soil samples with two kg weight were selected with two different soil textures and the solutions of soil were made with t...
متن کاملDechlorination of Polychlorinated Biphenyls in Real Waste Transformer Oil Using a Modified Household Microwave Oven in Presence of Sodium Hydroxide and Polyethylene Glycol
This research was carried out to assess the dechlorination and decomposition of Polychlorinated biphenyls (PCBs) in the real waste transformer oil using polyethylene glycol 1000/NaOH through a modified household microwave oven. To do so, the influence of polyethylene glycol (PEG) (1.5-7.5 g) and NaOH (0.3-1.5 g) under microwave (MW) power of 500 W on the dechlorination efficiency of PCBs, exist...
متن کاملDestruction and Dechlorination of Aroclor1254 in Real Waste Transformer Oil Using Microwave Irradiation, Microwave Absorbent and Reactive Materials
Background: In this research, the decomposition of Aroclor 1254, including 20 kinds of poly chlorinated biphenyls (PCBs), in real waste transformer oil in the presence of PEG1000, NaOH, zero valent iron and H2O by microwave (MW) irradiation was studied. Methods: Central composites design (CCD) method was carried out for experimental design. The effects of microwave power levels (200- 1000 W), ...
متن کاملEnergy and chemicals
Samples containing buprenorphine and derivatives were extracted using Agilent sample prep and analyzed by gas chromatography using an Agilent J&W HP-5 column. Published by Elsevier B. V. An empirical study on the selection of analytes and corresponding cutoffs for immunoassay and GC–MS in a two-step test strategy—buprenorphine example Analyst, 9, 1848-1856 (2009) Meng-Yan Wu et al. Tags HP-5, c...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2007