Specificity of sensitive assays of thyrotropin (TSH) used to screen for thyroid disease in hospitalized patients.
نویسندگان
چکیده
Thyrotropin (TSH) concentrations were measured in 1580 hospitalized patients and 109 normal persons. Using the mean +/- 3 SD limits of the log values for the controls (0.35-6.7 milli-int. units/L), the proportion of abnormal TSH results in the hospitalized patients was 17.2%. TSH was undetectable (less than 0.1 milli-int. unit/L) in 3.1% of patients, suggesting hyperthyroidism, and high (greater than 20 milli-int. units/L) in 1.6%, suggesting hypothyroidism. On follow-up of 329 patients, 62% with abnormal TSH (less than 0.35 or greater than 6.7 milli-int. units/L) and 38% with normal TSH concentrations, only 24% of those with undetectable TSH had thyroid disease: 36% of them were being treated with glucocorticoids and 40% had nonthyroidal illness (NTI). Although half the patients with TSH greater than 20 milli-int. units/L had thyroid disease, 45% of patients had high TSH values associated with NTI. TSH concentrations usually returned towards normal when patients' therapy with glucocorticoids was discontinued or they recovered from NTI. TSH test sensitivity appeared good when the mean +/- 3 SD limits of the reference population were used, i.e., no cases of hyper- or hypothyroidism, as identified by free thyroxin index (FT4I), were missed. However, TSH test specificity was inferior to that of the FT4I test (90.7% vs 92.3%), although specificity could be improved to 97.0% if the wider TSH reference limits of 0.1 to 20 milli-int. units/L were used--limits considered pathological if applied to outpatients. Evidently, different reference intervals for TSH are needed for hospitalized and nonhospitalized patients. We conclude that a "sensitive TSH assay" is not a cost-effective thyroid screening test for hospitalized patients as compared with the FT4I.
منابع مشابه
The measurement technologies of thyrotropin receptor antibodies from the past to the present
Thyroid Stimulating Hormone Receptor (TSH-R) autoantibodies are the main cause of Graveschr('39') disease and its external thyroid manifestations such as ophtalmopathy and dermatopathy. These antibodies are functionally different and are commonly called TSH receptor antibodies (TRAbs). In fact, TRAbs are a set of autoantibodies including TSHR-stimulating antibodies (TSAbs), TSHR- blocking antib...
متن کاملFour sensitive thyrotropin assays critically evaluated and compared.
We compared four sensitive procedures for thyrotropin (TSH)--Corning's Magic-Lite, ElectroNucleonics' Delfia, Baxter's Stratus, Nichols' Allegro--for their ability to completely discriminate TSH concentrations in sera in euthyroidism, hyperthyroidism, and hypothyroidism. We evaluated the analytical and clinical performance of these procedures according to previously published criteria. All proc...
متن کاملLinking medical needs and performance goals: clinical and laboratory perspectives on thyroid disease.
Diagnosis and management of thyroid disease benefit substantially from close interactions between clinicians and laboratorians. Clinicians desire reliable tests for diagnosis and management of both hypothyroidism and thyrotoxicosis. Traditionally, multitiered testing has been used, beginning with an index of free thyroxine concentration, which, if abnormal, has been followed by basal thyrotropi...
متن کاملGuidelines for diagnosis and monitoring of thyroid disease: nonthyroidal illness.
On the basis of low specificity, poor positive predictive value, and cost, there is at present no basis for routine assessment of thyroid function in acutely hospitalized patients, unless clinical features suggest the possibility of thyroid dysfunction, or a patient's background increases the likelihood of thyroid dysfunction. When used in severely ill patients, estimates of both thyroxine (T4)...
متن کاملUnstimulated high sensitive thyroglobulin measurement predicts outcome of differentiated thyroid carcinoma.
BACKGROUND Thyroglobulin (Tg) measurement following thyrotropin (TSH) stimulation is used in the follow-up of patients with differentiated thyroid carcinoma (DTC). However, high-sensitive assays allow accurate measurement of serum Tg even without TSH stimulation. Here, we prospectively evaluated the impact of unstimulated high-sensitive Tg measurement in early and long-term outcome of patients ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 33 8 شماره
صفحات -
تاریخ انتشار 1987