Commissioning of a mobile electron accelerator for intraoperative radiotherapy
نویسندگان
چکیده
Radiation performance characteristics of a dedicated intraoperative accelerator were determined to prepare the unit for clinical use. The linear accelerator uses standing wave X-band technology (wavelength approximately 3 centimeters) in order to minimize the mass of the accelerator. The injector design, smaller accelerator components, and low electron beam currents minimize radiation leakage. The unit may be used in a standard operating room without additional shielding. The mass of the accelerator gantry is 1250 Kg (weight approximately 2750 lbs) and the unit is transportable between operating rooms. Nominal electron energies are 4, 6, 9, and 12 MeV, and operate at selectable dose rates of 2.5 or 10 Gray per minute. D(max) depths in water for a 10 cm applicator are 0.7, 1.3, 1.7, and 2.0 for these energies, respectively. The depths of 80% dose are 1.2, 2.1, 3.1, and 3.9 cm, respectively. Absolute calibration using the American Association of Physicists in Medicine TG-51 protocol was performed for all electron energies using the 10 cm applicator. Applicator sizes ranged from 3 to 10 cm diameter for flat applicators, and 3 to 6 cm diameter for 30 degrees beveled applicators. Output factors were determined for all energies relative to the 10 cm flat applicator. Central axis depth dose profiles and isodose plots were determined for every applicator and energy combination. A quality assurance protocol, performed each day before patient treatment, was developed for output and energy constancy.
منابع مشابه
Commissioning the First Mobile Dedicated Accelerator for Intraoperative Electron Radiotherapy in Iran
Introduction: Intraoperative radiotherapy is a radiotherapy technique in which a high single fraction of radiation dose is delivered to the patient after surgery and Concurrent with anesthesia time. The most frequent method for IORT implementation is Intraoperative electron radiotherapy (IOERT), in which, some dedicated and high dose per pulse electron accelerators are employe...
متن کاملIntraoperative radiation therapy using mobile electron linear accelerators: report of AAPM Radiation Therapy Committee Task Group No. 72.
Intraoperative radiation therapy (IORT) has been customarily performed either in a shielded operating suite located in the operating room (OR) or in a shielded treatment room located within the Department of Radiation Oncology. In both cases, this cancer treatment modality uses stationary linear accelerators. With the development of new technology, mobile linear accelerators have recently becom...
متن کاملShielding assessment of a mobile electron accelerator for intra‐operative radiotherapy
A new approach to intraoperative radiotherapy makes use of a mobile electron linear accelerator delivering therapeutic radiation doses in an operating room suite. This unconventional technology has raised questions concerning protection for personnel and the necessity of shielding the adjoining areas. In this study, the leakage and scatter radiation from the mobile electron accelerator is measu...
متن کاملMonte Carlo Simulation of Electron Beams produced by LIAC Intraoperative Radiation Therapy Accelerator
Background: One of the main problems of dedicated IORT accelerators is to determine dosimetric characteristics of the electron beams. Monte Carlo simulation of IORT accelerator head and produced beam will be useful to improve the accuracy of beam dosimetry.Materials and Methods: Liac accelerator head was modeled using the BEAMnrcMonte Carlo simulation system. Phase-space files were generated at...
متن کاملشبیهسازی شتاب دهنده پرتو درمانی حین عمل LIAC به همراه اپلیکاتور شکل دهنده باریکه بااستفاده از محاسبات مونتکارلو
Intraoperative electron radiotherapy is one of the radiotherapy methods that delivers a high single fraction of radiation dose to the patient in one session during the surgery. Beam shaper applicator is one of the applicators that is recently employed with this radiotherapy method. This applicator has a considerable application in treatment of large tumors. In this study, the dosimetric charact...