Repeatability of central corneal thickness measurement with the Pentacam HR system

نویسندگان

  • Bruno Machado Fontes
  • Marcelo Palis Ventura
  • Renato Ambrósio
چکیده

Purpose: To assess the repeatability of central corneal thickness measurement at the geometrical center (Central Corneal Thickness CCT) given by the Pentacam High Resolution (HR) Comprehensive Eye Scanner (Oculus, Wetzlar, Germany) over time. Methods: Prospective, single center, observational study. Two separate CCT measurements were taken by the Pentacam corneal tomography exam (CTm) 3 to 12 months apart, and compared. Results: One hundred and sixteen eyes (n=116) of 62 health patients were included in this study. Average CCT in first and last visits was 541.6±37 μm and 543.6±36.9 μm respectively. Mean difference between both measurements was 9.2±6.4 μm, and there was no statistically significant difference in CCT measurement between visits, with good correlation between them (P = 0.057, r2 = 0,9209). Conclusion: Pentacam (HR) CTm gives repeatable CCT measurements over time.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry.

PURPOSE To assess the intraoperator repeatability and interoperator reproducibility of central corneal thickness measurements by the Pentacam Scheimpflug imaging system (Oculus) and the optical low-coherence reflectometer (OLCR) pachymeter (Haag-Streit) and to compare them with those of ultrasound (US) pachymetry. SETTING Assaf Harofe Medical Center Ophthalmology Outpatient Clinic, Zerifin, I...

متن کامل

Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices

Background. To estimate repeatability and comparability of central corneal thickness (CCT) and keratometry measurements obtained by four different devices in healthy eyes. Methods. Fifty-five healthy eyes from 55 volunteers were enrolled in this study. CCT (IOLMaster 700, Pentacam HR, and Cirrus HD-OCT) and keratometry readings (IOLMaster 700, Pentacam HR, and iDesign) were measured. For statis...

متن کامل

Comparison of Central Corneal Thickness Measured through Pentacam, OCTCasia2, and Orbscan IIz in Normal Individuals

Purpose: To compare the central corneal thickness measured through Pentacam, OCTCasia2, and Orbscan IIz in normal individuals. Methods: In this cross sectional comparative study, the researcher investigated the eyes of 70 normal candidates (140 eyes) with the mean age of 31.42 ± 7.70, referring to Dr. Sedaghat eye clinic for refractive surgery. For each person, Central Corneal Thickness (CCT) w...

متن کامل

Identifying more reliable parameters for the detection of change during the follow-up of mild to moderate keratoconus patients

Background Reaching a consensus on which parameters are most reliable at detecting progressive keratoconus patients with serial topography imaging is not evident. The aim of the study was to isolate the parameters best positioned to detect keratoconus progression using the Pentacam HR® measures based on the respective limits of repeatability and range of measurement. Method Using the Pentacam...

متن کامل

Central corneal thickness measurements with different imaging devices and ultrasound pachymetry.

PURPOSE The aim of this study was to compare 4 methods of central corneal thickness (CCT) measurements in terms of their agreement, repeatability, and measurement time. METHODS CCT was measured in 184 eyes of 92 healthy subjects by the same examiner. The methods used were as follows: noncontact specular microscopy (SM; Topcon SP-3000P; Topcon Corporation, Tokyo, Japan), Pentacam rotating Sche...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2012