[Scatter correction in myocardial thallium SPECT: needs for optimization of energy window settings in the energy window-based scatter correction techniques].
نویسندگان
چکیده
Accuracy and limitation of energy-window based scatter correction techniques have been evaluated for myocardial 201Tl SPECT by means of Monte Carlo simulation. In particular, projection view-dependency of energy distribution of the scattered photons was evaluated. Two geometrical configurations were simulated: namely a homogeneous cylindrical radioactivity located asymmetrically in a homogeneous cylindrical phantom, and a homogeneous ring radioactivity positioned at the myocardial region of a human thorax phantom. Energy spectra were recorded for each projection, and accuracy of the triple-energy window (TEW) method was then evaluation for both phantoms. The energy distribution of the scattered photons was apparently dependent on the projection view. TEW also demonstrated systematic overcorrection for the scatter because of multiple photo peaks around 80 KeV, and more importantly, the error was highly dependent on the projection view. The error reached to 35-38% for the view that is the closest to the 201Tl radioactivity (anterior view in case of the myocardial ring phantom), and was approximately 20% in the opposite view. This view-dependency of the error remained for other energy window settings, and was found to cause significant artifact in the reconstructed myocardial images, typically causing a defect in the anterior myocardial wall. Thus, this study demonstrated the need for optimizing the window settings for each projection view in all energy window-based scatter correction methods.
منابع مشابه
Simulation and patient studies of scatter correction in cardiac SPECT imaging
Introduction: Myocardial perfusion imaging is a nuclear medicine imaging method that is used to detect coronary artery diseases. One of the main sources of error in this imaging method is the detection of Compton scattered photons in the photopeak energy window used for data acquisition. This results in the degradation of the image contrast, and therefore decreases the...
متن کاملDetermination of the Energy Windows for the Triple Energy Window Scatter Correction Method in Gadolinium-159 Single Photon Emission Computed Tomography Using Monte Carlo Simulation
Introduction: In radionuclide imaging, object scatter is one of the major factors leading to image quality degradation. Therefore, the correction of scattered photons might have a great impact on improving the image quality. Regarding this, the present study aimed to determine the main and sub-energy windows for triple energy window (TEW) scatter correction method usin...
متن کاملThe Effect of FEW Scatter Correction Method in In-111 Imaging
Introduction: In nuclear medicine imaging, detection of the scattered photons along with the primary photons is one of the major problems that can lead to a decrease in ddiagnostic accuracy. Therefore, use of a suitable scatter correction method can help to improve of the image quality and increase of diagnostic accuracy. The aim of this study is evaluation of five energy wind...
متن کاملA new approach to scatter correction in SPECT images based on Klein_Nishina equation
Introduction: Scattered photon is one of the main defects that degrade the quality and quantitative accuracy of nuclear medicine images. Accurate estimation of scatter in projection data of SPECT is computationally extremely demanding for activity distribution in uniform and non-uniform dense media. Methods: The objective of this paper is to develop and validate a scatter correction technique ...
متن کاملEnergy window setting for optimum Tl-201 cardiac imaging [Persian]
Introduction: Poor sensitivity and poor signal to noise ratio because of low injected thallium dose and presence of scattered photons are the main problems in using thallium in scintigraphic imaging of the heart. Scattered photons are the main cause of degrading the contrast and resolution in SPECT imaging that result in error in quantification. Thallium decay is ve...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Kaku igaku. The Japanese journal of nuclear medicine
دوره 36 2 شماره
صفحات -
تاریخ انتشار 1999