Evaluation of Two Novel Integrated Stand-Alone Spacer Designs Compared with Anterior and Anterior-Posterior Single-Level Lumbar Fusion Techniques: An In Vitro Biomechanical Investigation
نویسندگان
چکیده
Study Design In vitro biomechanical investigation. Purpose To compare the biomechanics of integrated three-screw and four-screw anterior interbody spacer devices and traditional techniques for treatment of degenerative disc disease. Overview of Literature Biomechanical literature describes investigations of operative techniques and integrated devices with four dual-stacked, diverging interbody screws; four alternating, converging screws through a polyether-ether-ketone (PEEK) spacer; and four converging screws threaded within the PEEK spacer. Conflicting reports on the stability of stand-alone devices and the influence of device design on biomechanics warrant investigation. Methods Fourteen cadaveric lumbar spines were divided randomly into two equal groups (n=7). Each spine was tested intact, after discectomy (injured), and with PEEK interbody spacer alone (S), anterior lumbar plate and spacer (AP+S), bilateral pedicle screws and spacer (BPS+S), circumferential fixation with spacer and anterior lumbar plate supplemented with BPS, and three-screw (SA3s) or four-screw (SA4s) integrated spacers. Constructs were tested in flexion-extension (FE), lateral bending (LB), and axial rotation (AR). Researchers performed one-way analysis of variance and independent t-testing (p≤0.05). Results Instrumented constructs showed significantly decreased motion compared with intact except the spacer-alone construct in FE and AR (p≤0.05). SA3s showed significantly decreased range of motion (ROM) compared with AP+S in LB (p≤0.05) and comparable ROM in FE and AR. The three-screw design increased stability in FE and LB with no significant differences between integrated spacers or between integrated spacers and BPS+S in all loading modes. Conclusions Integrated spacers provided fixation statistically equivalent to traditional techniques. Comparison of three-screw and four-screw integrated anterior lumbar interbody fusion spacers revealed no significant differences, but the longer, larger-diameter interbody spacer with three-screw design increased stabilization in FE and LB; the diverging four-screw design showed marginal improvement during AR.
منابع مشابه
THE EFFECTS OF ANTERIOR DISCECTOMY AND INTERPOSITION AL IMPLANT UPON LUMBAR MOTION SEGMENT STABILITY
The cadaver spine motion segment behavior under torsional load was evaluated with the disc intact, with partial anterior discectomy and with spacer insertion. The results of this study explain how anterior lumbar discectomy and interbody fusion (ALIF) affects the torsional stability of the motion segment. The pseudarthrosis rate of the anterior lumbar discectomy and interbody fusion (ALIF)...
متن کاملComparison of Supplementary Posterior Fixations for Two-Level ALIF with Stand-Alone Cages
INTRODUCTION: Structural cages have been widely used in recent years to maintain motion segment height and stability in anterior lumbar interbody fusion (ALIF). High rate of clinical success and radiological fusion rate have been reported following single-level fusion using these cages. However, many biomechanical studies report insufficient stability, especially with extension loading. Posteri...
متن کاملComparative biomechanical evaluation a lumbar spacer with integrated screws
INTRODUCTION The goal of lumbar spinal fusion is to limit pain by decompression and removal of soft tissues, followed by the formation of a rigid bony union between adjacent vertebrae. The general consensus for a rigid fusion construct requires careful preparation of the intervertebral space and the creation of a rigid fusion construct. Anterior lumbar interbody fusion (ALIF) provides good acce...
متن کاملRevisiting Anterior Stand-Alone Fixation (ASAF) Devices for the Treatment of Single Level Lumbosacral Degenerative Disease
locking screw technology, which significantly augments the overall construct strength and rigidity. The goal of the anterior stabilized stand-alone device is to negate the need for posterior fixation by promoting stability of the implant via the locking screws. The benefit of creating an anterior stand-alone fixation device is that the spine surgeon can achieve adequate stabilization to ensure ...
متن کاملبررسی عوارض جراحی پولیواسکولیوز
This is a retrospective study of 83 patients with polioscoliosis, treated surgically in Shafa hospital (1970-1997), and the results are as Follow: implant failure: 15 cases (18%), pseudoar throsis: 12 cases (14.5%) deep infection: 4 cases (4.5%) superficial infection: 3 cases (3.6%) cast sore: 17 cases (20.5%) and loss of correction: 28 cases (33.7%). Implant failure was more common in Harr...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 11 شماره
صفحات -