Precision measurements of a nanoscale sample surface using an atomic force microscope (AFM) require a precise quantitative knowledge of the 3D tip shape. Blind tip reconstruction (BTR), established by Villarrubia, gives an outer bound with larger errors if the tip characterizer is not appropriate. In order to explore the errors of BTR, a series of simulation experiments based on a conical model were carried out. The results show that, to reconstruct the tip precisely, the cone angle of the tip characterizer must be smaller than that of the tip. Furthermore, the errors decrease as a function of the tip cone angle and increase linearly with the sample radius of curvature, irrespective of the tip radius of curvature. In particular, for sharp (20 nm radius) and blunt (80 nm radius) tips, the radius of curvature of the tip characterizer must be smaller than 5 nm. Based on these simulation results, a local error model of BTR was established. The maximum deviation between the errors derived from the model and the simulated experiments is 1.22 nm. Compared with the lateral resolution used in the above simulated experiments (4 nm/pixel), it is valid to ignore the deviations and consider the local error model of BTR is indeed in quantitative agreement with the simulation results. Finally, two simulated ideal structures are proposed here, together with their corresponding real samples. The simulation results show they are suitable for BTR.
منابع مشابه
Towards easy and reliable AFM tip shape determination using blind tip reconstruction.
Quantitative determination of the geometry of an atomic force microscope (AFM) probe tip is critical for robust measurements of the nanoscale properties of surfaces, including accurate measurement of sample features and quantification of tribological characteristics. Blind tip reconstruction, which determines tip shape from an AFM image scan without knowledge of tip or sample shape, was establi...
متن کاملAlgorithms for Scanned Probe Microscope Image Simulation, Surface Reconstruction, and Tip Estimation
To the extent that tips are not perfectly sharp, images produced by scanned probe microscopies (SPM) such as atomic force microscopy and scanning tunneling microscopy are only approximations of the specimen surface. Tip-induced distortions are significant whenever the specimen contains features with aspect ratios comparable to the tip's. Treatment of the tip-surface interaction as a simple geom...
متن کاملThe regularized blind tip reconstruction algorithm as a scanning probe microscopy tip metrology method
The problem of an accurate tip radius and shape characterization is very important for determination of surface mechanical and chemical properties on the basis of the scanning probe microscopy measurements. We think that the most favorable methods for this purpose are blind tip reconstruction methods, since they do not need any calibrated characterizers and might be performed on an ordinary SPM...
متن کاملInvestigation of Anisotropy Caused by Cylinder Applicator on Dose Distribution around Cs-137 Brachytherapy Source using MCNP4C Code
Introduction: Brachytherapy is a type of radiotherapy in which radioactive sources are used in proximity of tumors normally for treatment of malignancies in the head, prostate and cervix. Materials and Methods: The Cs-137 Selectron source is a low-dose-rate (LDR) brachytherapy source used in a remote afterloading system for treatment of different cancers. This system uses active and inactive sp...
متن کاملImage Simulation and Surface Reconstruction of Undercut Features in Atomic Force Microscopy
CD-AFMs (critical dimension atomic force microscopes) are instruments with servo-control of the tip in more than one direction. With appropriately “boot-shaped” or flared tips, such instruments can image vertical or even undercut features. As with any AFM, the image is a dilation of the sample shape with the tip shape. Accurate extraction of the CD requires a correction for the tip effect. Anal...