Compton Scatter and X-ray Crosstalk and the Use of Very Thin Intercrystal Septa in High-Resolution PET Detectors
نویسندگان
چکیده
To improve spatial resolution, positron emission tomography (PET) systems are being developed with finer detector elements. Unfortunately, using a smaller crystal size increases intercrystal Compton scatter and X-ray escape crosstalk, causing positioning errors that can lead to degradation of image contrast. We investigated the use of extremely thin ( 300 m) lead strips for passive shielding of this intercrystal crosstalk. Using annihilation gamma rays and small (2and 3-mm wide) Bismuth Germanate (BGO) crystal detectors in coincidence, crosstalk studies were performed with either two small adjacent crystals [(one-dimensional) (1-D)] or one crystal inside a volume of BGO [(two-dimensional) (2-D)]. The fraction of Compton scattered events from one crystal into an adjacent one was reduced, on average, by a factor of 3.2 (2.2) in the 1-D experiment and by a factor of 3.0 (2.1) in 2-D one, with a 300 (150)m-thick lead strip in between the crystals and a 300–700-keV energy window in both crystals. We could not measure a reduction in bismuth X-ray crosstalk with the use of lead septa due to the production of lead X-rays of similar energy. The full-width at half-maximum (FWHM) of the coincident point-spread function (CPSF) was not significantly different for the 1and 2-D studies, with or without the different septa in place. However, the FWTM was roughly 20% smaller with the 300m lead shielding in place. These results indicate that intercrystal crosstalk does not affect the positioning resolution at FWHM, but does affect the tails of the CPSF. Thus, without introducing any additional dead area, an insertion of very thin lead strips can reduce the extent of positioning errors. Reducing the intercrystal crosstalk in a high-resolution PET detector array could potentially improve tomographic image contrast in situations where intercrystal crosstalk plays a significant role in event mispositioning.
منابع مشابه
COMPTON SCATTER AND X-RAY CROSSTALK AND THE USE OF VERY THIN INTER-CRYSTAL SEPTA IN HIGH RESOLUTION - Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
To improve spatial resolution, PET systems are being developed with finer detector elements. Unfortunately, using smaller crystal sizes increases inter-crystal Compton scatter and bismuth x-ray and (to a lesser degree) electron escape crosstalk, causing positioning errors that lead to degradation of image contrast. We investigated the use of extremely thin (I 300 ym) lead strips for passive shi...
متن کاملAssessment of X-Ray Crosstalk in a Computed Tomography Scanner with Small Detector Elements Using Monte Carlo Method
Introduction: Crosstalk is a leakage of X-ray or light produced in a matrix of X-ray detectors or array of photodiodes in one element to other elements affecting on image contrast and spatial resolution. In this study, we assessed X-ray crosstalk in a computed tomography (CT) scanner with small detector elements to estimate the effect of various parameters such as X-ray tube voltage, detector e...
متن کاملThe Influence of Crystal Size and Material on Intercrystal Scattering and Parallax in PET Block Detectors: A Monte Carlo Study
Introduction: In this study, we utilized the MCNP4C Monte Carlo code to quantitatively evaluate the influence of crystal size and material on intercrystal scatter and parallax effects. Materials and Methods: For each of the 5 selected crystals (BGO, LSO, LYSO, LuAP, GSO), transport of 511 keV photons originating from a point source and incident on the central cry...
متن کاملUnfolding X-ray spectrum in the diagnostic range using the Monte Carlo Code MCNP5
Introduction: Unfolding X-ray spectrum is a powerful tool for quality control of X-ray tubes. Generally, the acquisition of the X-ray spectrum in diagnostic radiology departments is complicated and difficult due to high photon flux. Measurement of x ray spectra using radiation detectors could not be performed accurately, because of the pulse pile up. Therefore, indirect methods...
متن کاملF eb 2 00 4 Measurements of Compton Scattered Transition Radiation at High Lorentz Factors
X-ray transition radiation can be used to measure the Lorentz factor of relativistic particles. Standard transition radiation detectors (TRDs) typically incorporate thin plastic foil, foam, or fiber radiators and gas-filled x-ray detectors, and are sensitive up to γ ∼ 10. To reach Lorentz factors up to γ ∼ 10, thicker, denser radiators can be used, which consequently produce x-rays of harder en...