K3 Endo, ProTaper, and ProFile systems: breakage and distortion in severely curved roots of molars.

نویسندگان

  • Matthew T Ankrum
  • Gary R Hartwell
  • John E Truitt
چکیده

It was the aim of this study to investigate the incidence of file breakage and distortion when the ProTaper, K3 Endo, and ProFile systems were used to instrument canals in the severely curved roots of extracted molars. Forty-five roots of extracted mandibular and maxillary molars with curvatures between 40 and 75 degrees were chosen for this study. The canals in group 1 were instrumented with the ProFile system. Roots in group 2 were instrumented with the ProTaper system, and those in group 3 were instrumented with the K3 Endo system. The three systems were used according to the manufacturers' instructions. The proportion of files distorted was 15.3% for the ProFile group, 2.4% for the ProTaper group, and 8.3% for the K3 Endo group. There was a statistically significant difference between the ProFile and ProTaper groups (p=0.0079). The percentage of broken files was 1.7% for the ProFile group, 6.0% for the ProTaper group, and 2.1% for the K3 Endo group. No statistically significant differences were found between these three groups (p=0.4243). The results of this study showed that these three rotary tapered systems were not significantly different with regard to breakage. There were significantly more distorted files in the ProFile group when compared with the ProTaper group. With regard to distortion, there was no significant difference between the ProTaper and K3 Endo and the ProFile and K3 Endo groups.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Relative efficacy of three Ni-Ti file systems used by undergraduates

The purpose of this study was to compare and evaluate the shaping ability of the three different Ni-Ti file systems used by undergraduate students. Fifty undergraduate students prepared 150 simulated curved root canals in resin blocks with three Ni-Ti file systems ProFile (PF), Manual ProTaper (MPT), Rotary ProTaper (RPT). Every student prepared 3 simulated root canals with each system respecti...

متن کامل

EVALUATION OF APICAL DEVIATION IN ROOT CANALS INSTRUMENTED WITH K3 AND ProTaper SYSTEMS

OBJECTIVES This study evaluated the apical deviation of curved root canals instrumented with K3 and ProTaper systems. MATERIAL AND METHODS Twenty root canals of human maxillary and mandibular first molars were employed, which were divided into 2 groups: group A (10 teeth) was instrumented with the K3 system, and group B (10 teeth) with the ProTaper system. Evaluation of deviation was performe...

متن کامل

A Comparative Study on the Shaping Ability of K3, Profile and Protaper Instruments in Simulated Curved Root Canals

INTRODUCTION The objective of this study was to compare the shaping ability of three rotary filing systems; constant taper K3 instruments, constant taper ProFile instruments and progressive taper ProTaper rotary instruments in clear resin blocks with simulated curved root canals. MATERIALS AND METHODS Forty five resin blocks were divided into three groups. Group A preparation was conducted wi...

متن کامل

In vitro Comparison of Apical Debris Extrusion Using Rotary and Reciprocating Systems in Severely Curved Root Canals

INTRODUCTION The aim of this in vitro study was to compare the amount of apically extruded debris after root canal preparation using rotary and reciprocating systems in severely curved root canals. METHODS AND MATERIALS Thirty six extracted human mandibular first molars with 25-35° curvature in their mesiobuccal (MB) canal (according to Schneider's method) were cleaned and shaped with ProTape...

متن کامل

In-vitro Comparison of NiTi Tee, RaCe and Protaper Instruments in Dentinal Crack Formation

Background and Aim:Root canal preparation with rotary instruments may cause dentinal cracks leading to tooth fracture. The aim of this study was to compare three different rotary systems ProTaper, RaCe and Niti Tee on formation of dentinal cracks following root canal preparation. Materials and Methods: In this experimental study, 50 extracted mandibular first molars were selected. Teeth having ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of endodontics

دوره 30 4  شماره 

صفحات  -

تاریخ انتشار 2004