The analyzation and clinical evaluation of ECLIA and CMIA in the detection of Treponema pallidum

نویسندگان

  • Jiansuo Zhou
  • Yongming Liang
  • Jie Zhang
  • Liyan Cui
چکیده

To evaluate the clinical application value of electrochemiluminescence immunoassay analyzer (ECLIA) and chemiluminescent magnetic microparticle immunoassay (CMIA) in the detection of Treponema pallidum (TP).A total of 1225 patients in Peking University Third Hospital was enrolled from June 2014 to October 2014. ECLIA and CMIA were applied to detect the serum anti-TP. The positive rate was analyzed. RIBA was taken as a golden standard to evaluate the sensitivity, the specificity, the positive predictive value, the negative predictive value, and the accuracy of ECLIA and CMIA. A correlation analysis between 2 assays was conducted, and that between assay and RIBA. We also evaluate the clinical value of TPPA in the detection of T pallidum.The positive rate of CMIA and ECLIA is 10.63% and 9.89%, respectively, showing no statistically significant difference (P > .05). For CMIA, ECLIA, and TPPA, the sensitivity is 99.16%, 99.16%, and 99.16%, the specificity is 98.99%, 99.82%, and 100%, the positive predictive value is 91.47%, 98.33%, and 100%, the negative predictive value is 99.91%, 99.91%, and 99.91%, the coincidence rate is 99.01% (Kappa = 0.895), 99.75% (Kappa = 0.997), and 99.92% (Kappa = 0.998), respectively.The result shows high correlation between ECLIA and CMIA. Both have high sensitivity and specificity and can be used as screening tests for the diagnosis of T pallidum in common condition.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Importance of extending the use of polymerase chain reaction in the diagnosis of venereal syphilis in a blood transfusion center in Burkina Faso, West Africa

INTRODUCTION Due to the existence of a variety of types of non-venereal syphilis caused by the related T. pallidum, regular serological testing such as Rapid Plasma Reagin (RPR) and Chemiluminescent Microparticle Immunoassay Technique (CMIA) are often unable to differentiate venereal syphilis from the non- venereal one, hence, the interest in the use of molecular biology testing for a confirmat...

متن کامل

Serologic diagnosis of syphilis: comparison of different diagnostic methods.

Diagnostic and therapeutic approaches in syphilis show wide variation. The use of only one type of serologic test is insufficient for diagnosis. However, current international recommendations cannot be applied due to various reasons (cost, availability, etc.). The aim of the study was to review serologic data of syphilis patients to determine diagnostic performance of three different methods. I...

متن کامل

Validation of reverse sequence screening for syphilis.

Syphilis, which is caused by Treponema pallidum, is a chronic bacterial infection that remains a public health concern worldwide. Serologic testing is the method most often used, as the bacterium cannot be cultured. Syphilis serological tests are divided into nontreponemal and treponemal tests, and neither is sufficient alone for diagnosis, as each type of test has limitations, including the po...

متن کامل

Preliminary evaluation of an immunochromatographic strip test for specific Treponema pallidum antibodies.

We evaluated a prototype immunochromatographic strip (ICS) test for qualitative detection of Treponema pallidum antibodies in 353 sera from 157 patients. For sera from 43 syphilis patients, the ICSs were reactive, while for sera from 114 patients without syphilis, including 22 with biologically false-positive Rapid Plasma Reagin test results, the ICSs were nonreactive. The ICS test may expand t...

متن کامل

Comparison of Traditional and Reverse Syphilis Screening Algorithms in Medical Health Checkups

BACKGROUND The syphilis diagnostic algorithms applied in different countries vary significantly depending on the local syphilis epidemiology and other considerations, including the expected workload, the need for automation in the laboratory and budget factors. This study was performed to investigate the efficacy of traditional and reverse syphilis diagnostic algorithms during general health ch...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 96  شماره 

صفحات  -

تاریخ انتشار 2017