Tissue reaction to implants of different metals: a study using guide wires in cannulated screws.

نویسندگان

  • D M Devine
  • M Leitner
  • S M Perren
  • L P Boure
  • S G Pearce
چکیده

Cannulated screws, along with guide wires, are typically used for surgical fracture treatment in cancellous bone. Breakage or bending deformation of the guide wire is a clinical concern. Mechanically superior guide wires made of Co-Cr alloys such as MP35N and L605 may reduce the occurrence of mechanical failures when used in combination with conventional (316L stainless steel) cannulated screws. However the possibility of galvanic or crevice corrosion and adverse tissue reaction, exists when using dissimilar materials, particularly in the event that a guide wire breaks, and remains in situ. Therefore, we designed an experiment to determine the tissue reaction to such an in vivo environment. Implant devices were designed to replicate a clinical situation where dissimilar metals can form a galvanic couple. Histological and SEM analyses were used to evaluate tissue response and corrosion of the implants. In this experiment, no adverse in vivo effects were detected from the use of dissimilar materials in a model of a broken guide wire in a cannulated screw.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Corrosion and Tissue Reaction to MP35N, L605 & 316L Guide Wires in Combination with a Conventional 316L Cannulated Stainless Steel Screw

INTRODUCTION: Cannulated screws, along with guide wires, are commonly used for a variety of orthopedic indications. Typically they are used for fractures in cancellous regions. Breakage or bending deformation of the guide wire is a clinical concern. Therefore it would be advantageous to use Co-Cr alloys such as MP35N and L605 in these applications as the occurrence of breakages and bendings wou...

متن کامل

Implant Removal Matrix for the Upper Extremity Orthopedic Surgeon

Orthopedic implant removal is a commonly performed procedure. While implant removal can be associated withimproved symptoms, risks of the surgery are notable. Stripped screws, broken and retained hardware, and morbidityassociated with soft tissue compromise during difficult removal are all common. Familiarity with the instruments iscritical to procedure success. The purpose of...

متن کامل

Intrapelvic Protrusion of a Broken Guide Wire Fragment during Fixation of a Femoral Neck Fracture

  During fixation of a femoral neck fracture in a 23-year-old male patient with cannulated screws, a broken guide wire fragment inadvertently advanced through the hip joint and protruded into the pelvis. A second surgical approach was needed to remove the broken fragment from the pelvis. Awareness of such a potentially devastating complication will make surgeons more cautious during implementa...

متن کامل

Letter to the Editor: Cannulated versus non-cannulated cancellous screw fixation for femoral neck fractures: a synthetic bone biomechanical study.

PURPOSE To compare the compressive strength of cannulated versus non-cannulated cancellous screws for undisplaced femoral neck fractures in synthetic bones. METHODS 18 synthetic proximal femurs simulating an AO B1 valgus impacted femoral neck fracture in osteoporotic bone were used. The fracture angles were between 55º and 63º (Pauwels grade 2). Fixation was made using 6.5-mm non-cannulated s...

متن کامل

A Novel Surgical Technique for Removing Buried Cannulated Screws Using a Guidewire and Countersink: A Report of Two Cases

Removal of metal implants is a common procedure that is performed for a variety of indications. However, problems such as a buried screw head may occasionally arise and render hardware removal difficult or even impossible. The problem is further compounded when the initial screw was inserted percutaneously or via a minimally-invasive (MIS) technique. In the present paper, we introduce a novel, ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • European cells & materials

دوره 18  شماره 

صفحات  -

تاریخ انتشار 2009