Preparation and use of serum-based material as control and calibrator in evaluating ion-selective electrodes for calcium.

نویسندگان

  • J Toffaletti
  • K M Lee
چکیده

We analyzed 93 sera and 81 whole-blood samples, plus several aqueous materials and one serum-based material as controls, with two analyzers for ionized calcium (Radiometer ICA 1 and AVL 980). On 16 days, comparisons of results for patients' samples were good, with nearly all between-instrument differences (delta Ca2+) of samples being within 0.04 mmol/L. On 12 days, comparisons were mediocre, with the delta Ca2+ of samples usually 0.04 to 0.10 mmol/L. The control material that most consistently indicated the direction and magnitude of the delta Ca2+ of patients' samples was the serum-based control. Adjustment of results for samples from patients, based on the delta Ca2+ of the serum control on the same day, substantially improved comparisons between instruments. Our findings suggest that the use of a serum-based calibrator provides the most valid comparison between different calcium-ion analyzers and reliably indicates when electrode replacement is needed.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A readily available source of serum for use as a precision quality-control material for ionized calcium measurement by ion-selective electrodes

A readily available source of serum ft r use as a precision quality-control material for ionized calcium measurement by ion-selective electrodes A simple method for preparing human serum for use as a precision quality-control for ionized calcium using an ion-selective electrode has been described earlier [1-1. Human citrate dextrose plasma is readily available from outdated whole blood from blo...

متن کامل

Development of novel potentiometric sensors for determination of lidocaine hydrochloride in pharmaceutical preparations, serum and urine samples

This article is focused on the determination of lidocaine hydrochloride as a local anaesthetic drug. A potentiometric method based on modified screen-printed and modified carbon paste ion-selective electrodes was described for the determination of lidocaine hydrochloride in different pharmaceutical preparations. It was based on potentiometric titration of lidocaine hydrochloride using modified ...

متن کامل

Development of novel potentiometric sensors for determination of lidocaine hydrochloride in pharmaceutical preparations, serum and urine samples

This article is focused on the determination of lidocaine hydrochloride as a local anaesthetic drug. A potentiometric method based on modified screen-printed and modified carbon paste ion-selective electrodes was described for the determination of lidocaine hydrochloride in different pharmaceutical preparations. It was based on potentiometric titration of lidocaine hydrochloride using modified ...

متن کامل

Potassium-Selective Membrane Sensor Based on Two Recently Synthesized Poly(oxyethylene) Derivatives

Two poly(oxyethylene) derivatives 1.10-bis(2'-benzoic acid)-1,4,7,1-tetradecane (L1) and 1,7-bis(2'-benzoic acid)-1,4,7-trioxaheptane (L2) were used as suitable ionophores for the preparation of potassium-selective PVC-membrane electrodes. The resulting electrodes based on L1 and L2 exhibit linear responses to the activity of K+ ions within the concentration ranges of 3.1<em...

متن کامل

Comparison of the vanadyl-selective electrodes prepared by sol–gel and PVC membrane techniques

Sol-gel and polymeric membrane electrodes, based on thiacalix[4]arene as a neutral carrier, weresuccessfully developed for the detection of VO2+ in aqueous solutions. The sol-gel and PVCmembrane electrodes exhibited linear response with Nernstian slopes of 29.3 and 28.4 mV perdecade, respectively, within the vanadyl ion concentration ranges from 1.0 × 10-6 to 1.0 × 10-1mol dm-3 and from 1.0 × 1...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 31 8  شماره 

صفحات  -

تاریخ انتشار 1985