Wear simulation of a polyethylene-on-metal cervical total disc replacement under different concentrations of bovine serum lubricant.

نویسندگان

  • P J Hyde
  • J Fisher
  • R M Hall
چکیده

Metal-on-polyethylene total disc replacements have been an alternative to spinal fusion in the lumbar spine under certain indications for more than a decade. Recently, cervical total disc replacement has also become an alternative to cervical fusion. Knowledge acquired from years of in vitro simulator studies on other joint replacements has highlighted the risks associated with premature wear due to unforeseen adverse clinical conditions and the effect of particulate debris on surrounding natural tissues. Having no evidence of the type and composition of the lubricating fluid that will result after spinal arthroplasty, a study on the effects of lubricant serum concentration was undertaken. The wear rate was shown to be inversely proportional to protein content of the serum over a range of 50%-3% bovine serum to water concentration.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Are PEEK-on-Ceramic Bearings an Option for Total Disc Arthroplasty? An In Vitro Tribology Study

BACKGROUND Most contemporary total disc replacements (TDRs) use conventional orthopaedic bearing couples such as ultrahigh-molecular-weight polyethylene (polyethylene) and cobalt-chromium (CoCr). Cervical total disc replacements incorporating polyetheretherketone (PEEK) bearings (specifically PEEK-on-PEEK bearings) have been previously investigated, but little is known about PEEK-on-ceramic bea...

متن کامل

Increased total knee arthroplasty ultra-high molecular weight polyethylene wear using a clinically relevant hyaluronic acid simulator lubricant.

In this study, osteoarthritic and periprosthetic synovial fluid samples were rheologically and biochemically compared to develop a hyaluronic acid (HA) supplemented bovine serum (BS) lubricant that mimicked the properties of human joint synovial fluid. The effect of this BS + HA lubricant (50 per cent bovine calf serum + 1.5 g/l HA) on the wear rate of ultra-high molecular weight polyethylene (...

متن کامل

Cervical Disc Prosthesis Polyethylene Wear Following The ISO Cervical Test

Introduction: Cervical total disc replacement (TDR) procedures are an option for treatment of cervical spinal disc degeneration. The goal of TDR replacement is to replace the painful and degenerated TDR with a new bearing unit to enable pain free motion. A young and active patient population demands a proven durable replacement bearing surface to accomplish this goal. The wear of that surface m...

متن کامل

Albumin Protein Cleavage Affects the Wear and Friction of Ultra-High Molecular Weight Polyethylene

It is well established that the total protein concentration and albumin-to-globulin ratio influence the wear of ultra-high molecular polyethylene (UHMWPE, "polyethylene") in joint prostheses. A factor on wear not yet studied, but of possible clinical relevance, is protein cleavage. Such cleavage is expected in the presence of an inflammatory response and as a result of wear processes at the art...

متن کامل

Effect of lubricant composition on the fatigue properties of ultra-high molecular weight polyethylene for total knee replacement.

Ultrahigh molecular weight polyethylene (UHMWPE) fatigue is a critical factor affecting the longevity of total knee replacement (TKR) bearings. With the increased need for laboratory studies to mimic near in vivo conditions for accurate characterization of material performance, the present study investigated the role of hyaluronic acid (HA) in testing lubricant on the crack growth response of U...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine

دوره 230 5  شماره 

صفحات  -

تاریخ انتشار 2016