Numerical Study and Ex Vivo Assessment of HIFU Treatment Time Reduction through Optimization of Focal Point Trajectory
نویسندگان
چکیده
Treatment time reduction is a key issue to expand the use of high intensity focused ultrasound (HIFU) surgery, especially for benign pathologies. This study aims at quantitatively assessing the potential reduction of the treatment time arising from moving the focal point during long pulses. In this context, the optimization of the focal point trajectory is crucial to achieve a uniform thermal dose repartition and avoid boiling. At first, a numerical optimization algorithm was used to generate efficient trajectories. Thermal conduction was simulated in 3D with a finite difference code and damages to the tissue were modeled using the thermal dose formula. Given an initial trajectory, the thermal dose field was first computed, then, making use of Pontryagin’s maximum principle, the trajectory was iteratively refined. Several initial trajectories were tested. Then, an ex vivo study was conducted in order to validate the efficicency of the resulting optimized strategies. Single pulses were performed at 3MHz on fresh veal liver samples with an Echopulse and the size of each unitary lesion was assessed by cutting each sample along three orthogonal planes and measuring the dimension of the whitened area based on photographs. We propose a promising approach to significantly shorten HIFU treatment time: the numerical optimization algorithm was shown to provide a reliable insight on trajectories that can improve treatment strategies. The model must now be improved in order to take in vivo conditions into account and extensively validated.
منابع مشابه
Numerical Study for Optimizing Parameters of High-Intensity Focused Ultrasound-Induced Thermal Field during Liver Tumor Ablation: HIFU Simulator
Introduction High intensity focused ultrasound (HIFU) is considered a noninvasive and effective technique for tumor ablation. Frequency and acoustic power are the most effective parameters for temperature distribution and the extent of tissue damage. The aim of this study was to optimize the operating transducer parameters such as frequency and input power in order to acquire suitable temperatu...
متن کاملبررسی انکسار پرتوهای فراصوت در هایفوتراپی و بهینه سازی عمق کانونی مبدل فراصوت جهت افزایش دقت درمان
Background and purpose: Nowadays High-intensity focused ultrasound (HIFU) as non-ionizing radiation is used for cancer treatment. Ultrasound beams when crossing the border of two environments are refracted causing decrease in the beam focusing within a tumor. Also, it may induce damage to normal tissues. In this study, we evaluated the focal depth shift induced by ultrasonic beam refraction. ...
متن کاملAssessment of the Performance of Clustering Algorithms in the Extraction of Similar Trajectories
In recent years, the tremendous and increasing growth of spatial trajectory data and the necessity of processing and extraction of useful information and meaningful patterns have led to the fact that many researchers have been attracted to the field of spatio-temporal trajectory clustering. The process and analysis of these trajectories have resulted in the extraction of useful information whic...
متن کاملHigh intensity focused ultrasound (HIFU) focal spot localization using harmonic motion imaging (HMI).
Several ultrasound-based imaging modalities have been proposed for image guidance and monitoring of high-intensity focused ultrasound (HIFU) treatment. However, accurate localization and characterization of the effective region of treatment (focal spot) remain important obstacles in the clinical implementation of HIFU ablation. Harmonic motion imaging for focused ultrasound (HMIFU) is a HIFU mo...
متن کاملOptimization of volumetric MR-guided high-intensity focused ulatrsound ablations in moving organs
Introduction In this work, a method for the optimization of volumetric MR-guided high intensity focused ultrasound (HIFU) ablations is presented. While commonly MRguidance of thermal therapies is based on temperature control the on-line assessment of the thermal dose has been suggested to determine the therapy endpoint [1]. Therefore, we propose to monitor the thermal dose [2] of the treated ti...