Double Spin-Echo Spiral Chemical Shift Imaging for Rapid Metabolic Imaging of Hyperpolarized [1-C]-Pyruvate
نویسندگان
چکیده
Introduction Rapid chemical shift imaging (CSI) techniques are of considerable interest for hyperpolarized C metabolic imaging [1,2]. Undersampled spiral CSI (spCSI) using a free induction decay (FID) acquisition mode has been applied to real-time metabolic imaging with acquisition times of less than 1 s per slice [3]. The FID acquisition necessitates phase correction during reconstruction for accurate spectral quantitation as the metabolite signals cannot be separated in magnitude mode. Phase correction can be challenging, especially with contributions from aliased out-of-phase peaks. However, when the full spin echo is acquired, the magnitude operation does not increase the linewidth, hence eliminating the need for phase correction. This work extends the spiral CSI sequence to incorporate a double spin-echo (DSE) [4] using a pair of adiabatic refocusing pulses. Methods All measurements were performed on a GE 3T MR scanner equipped with self-shielded gradients (40 mT/m, 150 mT/m/ms). A custom-built dual-tuned (H/C) quadrature coil (dia=80 mm, length=90 mm) was used for both RF excitation and signal reception. Healthy male Wistar rats were anesthetized with 1-3% isoflurane in oxygen (~1.5 l/min). The rats were injected in a tail vein with 2.5-3 ml of 80-mM solution of [1-C] pyruvate that was hyperpolarized using HyperSense DNP (15-20% liquid state polarization). 2D C data were acquired using the DSE-spCSI sequence with sliceselective excitation followed by 2 adiabatic refocusing pulses for a double spin-echo (pulses described in [4]) and a spiral readout trajectory that encoded the spectral and both spatial dimensions. The spiral waveforms were designed for FOV=80x80 mm and nominal 5x5 mm in-plane resolution, with 3 spatial interleaves and spectral width of 276.2 Hz. The TE was 151 ms for DSE-spCSI and 3 ms for FID-spCSI. The DSE acquisition was compared to FID in a single time-point measurement of a 10 mm axial slice through the kidneys. Imaging was started 20 s after the injection of hyperpolarized pyruvate. A variable-flip-angle scheme [5] of 35o, 45o, and 90o was used for the three spiral interleaves to account for depletion of the longitudinal magnetization with multiple excitations and excite the same amount of transverse magnetization for each acquisition. Longitudinal relaxation and metabolic turnover can be neglected for the short time duration between the excitations (125 ms for FID and 220 ms for DSE). The data were apodized with 15 Hz Gaussian line broadening in the time dimension and reconstructed as described in [3]. Metabolic maps for pyruvate (Pyr), lactate (Lac), alanine (Ala) and bicarbonate (Bic) were calculated by integrating the signal within 20 Hz (10 Hz for Bic) around each peak in absorption mode for FID and magnitude mode for DSE. The DSE-spCSI sequence was also used in dynamic imaging experiments and compared to FID-spCSI, for hyperpolarized Pyr solution in vitro and in vivo. A constant 5o flip angle excitation was used and images were acquired every 3 s, starting at the time of injection. Another DSE acquisition was done with the acquisition starting at the end of injection (15 s delay). Results and Discussion Figure 1 shows the metabolic images of Pyr, Lac, Ala and Bic superimposed onto a H single-shot FSE image (2-mm slice) for anatomical reference. The spectra (magnitude spectra for DSE and absorption spectra for FID) from a voxel in the kidney are also shown. The DSE-spCSI exhibits high quality spectra, with the magnitude mode linewidth being comparable to absorption mode FID-spCSI spectra. The Bic signal in particular is obscured in the FID acquisition due to the contribution from the out-of-phase aliased Pyr peak, but can easily be distinguished in the DSE magnitude spectrum. Automatic phase-correction of the FID spectra failed for Bic due to its proximity to the Pyr peak, and the spectrum from the kidney voxel was manually phase-corrected to obtain the Bic peak shown here. An added benefit of the DSE sequence is that it suppresses the signal in blood vessels, as crusher gradients around the refocusing RF pulses attenuate the magnetization from flowing spins. In the FID, the high Pyr signal in the vessels can contaminate the signal from the kidney at the resolution used here. To verify that the DSE-spCSI preserves signal over multiple excitations, the T1 of hyperpolarized Pyr solution in a syringe was measured. The T1 of 59 sec observed with DSE-spCSI was similar to the 60 sec measured with slice-selective FID acquisition. The time courses of the mean Pyr and Lac signals (corrected for polarization level) from an ROI in the right kidney are shown in Fig. 2. The DSE showed lower signal and decayed slightly faster than the FID, as the refocusing pulses spoil any magnetization passing through the transition bands/fringe field at the edges of the RF coil. This is a significant loss when there is Pyr flowing through anatomy lying near the coil edges. Delaying the acquisition until the end of injection may provide some gain in this case. Conclusion This work demonstrates the use of DSE-spCSI for fast metabolic imaging, attaining high quality spectra without the need for phase correction along with the advantage of attenuating signal in the vessels. Care must be taken for dynamic imaging to ensure that the spins remain within the RF coil volume.
منابع مشابه
3D cardiac Chemical Shift Imaging of [1-13C] hyperpolarized acetate and pyruvate in pigs
Background 13C Dynamic Nuclear Polarization (DNP) with rapid dissolution together with Magnetic Resonance Chemical Shift Imaging (CSI) have been used for non-invasive real-time metabolic assessment in cardiac experimental models on a clinical 3T scanner. Here, we report an in vivo comparison of hyperpolarized [1-13C] pyruvate and [1-13C] acetate perfusion and metabolism: a method based on a 3D ...
متن کاملMultiband excitation pulses for hyperpolarized 13C dynamic chemical-shift imaging.
Hyperpolarized 13C offers high signal-to-noise ratios for imaging metabolic activity in vivo, but care must be taken when designing pulse sequences because the magnetization cannot be recovered once it has decayed. It has a short lifetime, on the order of minutes, and gets used up by each RF excitation. In this paper, we present a new dynamic chemical-shift imaging method that uses specialized ...
متن کاملIn vivo single-shot 13C spectroscopic imaging of hyperpolarized metabolites by spatiotemporal encoding.
Hyperpolarized metabolic imaging is a growing field that has provided a new tool for analyzing metabolism, particularly in cancer. Given the short life times of the hyperpolarized signal, fast and effective spectroscopic imaging methods compatible with dynamic metabolic characterizations are necessary. Several approaches have been customized for hyperpolarized (13)C MRI, including CSI with a ce...
متن کاملMulti-resonance 3d Spin-echo Epi with Chemical Separation for Fast Hyperpolarized C Mri
Target Audience Hyperpolarized (HP) C MRI researchers. Purpose In hyperpolarized C experiments, the data acquisition window is fundamentally limited by rapid metabolism and T1 relaxation of the prepolarized spins. This constraint necessitates the adoptation of fast imaging schemes such as echo planar imaging (EPI) [1]. However, the application of EPI to HP C experiments results in images with c...
متن کاملMulti-band frequency encoding method for metabolic imaging with hyperpolarized [1-(13)C]pyruvate.
A new method was developed for simultaneous spatial localization and spectral separation of multiple compounds based on a single echo, by designing the acquisition to place individual compounds in separate frequency encoding bands. This method was specially designed for rapid and robust metabolic imaging of hyperpolarized (13)C substrates and their metabolic products, and was investigated in ph...